FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 1920933 · Received December 6, 2010

Report

Report Number
1824206-2010-11401
Event Type
Malfunction
Date Received
December 6, 2010
Date of Event
November 10, 2010
Report Date
November 10, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THAT THE AC POWER CORD PLUG HAD A LOOSE GROUND PIN. HE REPLACED THE AC PLUG WHICH RESOLVED THE HIGH GROUND RESISTANCE READING.

Description of Event or Problem · 1

INFO RECEIVED INDICATES, THE GROUND RESISTANCE READING WAS HIGH WHILE DOING AN ELECTRICAL SAFETY TEST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1604

Patients

Seq Age Sex Outcome Treatment
1