Report Number
2649622-2010-12349
Event Type
Injury
Date Received
December 13, 2010
Date of Event
August 26, 2010
Report Date
February 27, 2024
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
K902002
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. CORRECTED DATA: (DATE OF DEATH TO NOT APPLICABLE, PATIENT OUTCOME CHECKED HOSPITALIZATION).

Description of Event or Problem · 1

IT WAS REPORTED THAT DEVICE WAS INTERROGATED AND BATTERY VOLTAGE OF 2.62V WITH THE ERI FLAG NOT SET. THERE WAS CONCERN THAT DEVICE ONLY LASTED 3.5 YEARS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR LEAD WAS UNDERSENSING. THE RIGHT VENTRICULAR LEAD WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED THAT THE LEFT VENTRICULAR EPICARDIAL LEAD HAD INTERMITTENT CAPTURE. THE LEFT VENTRICULAR EPICARDIAL LEAD WAS CAPPED, AND ANOTHER EXISTING LEFT VENTRICULAR EPICARDIAL LEAD WAS CONNECTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SCREW-IN IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5071 ASKU

Patients

Seq Age Sex Outcome Treatment
1 60 YR Male Hospitalization| R 5071-1 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD| 5071-1 IMPLANTABLE PACING LEAD