FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 1920856 · Received December 13, 2010

Report

Report Number
2649622-2010-12345
Event Type
Injury
Date Received
December 13, 2010
Date of Event
August 24, 2010
Report Date
February 27, 2024
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DISTAL CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING AND A COSMETIC DEPRESSION. THE LEAD WAS STRETCHED AND APPEARED TO HAVE EXPLANT DAMAGE. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED) AND BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE INITIAL PROCEDURE, THE 4193 LV LEAD COULD CAPTURE VIA THE ANALYZER BUT NOT AFTER HOOKING UP TO THE DEVICE. IN A LATER PROCEDURE, THE 4193 LEAD WAS EXPLANTED BECAUSE OF OBSERVED HIGH THRESHOLD RESULTING IN NO PACING CAPTURE. AN IMPLANT ATTEMPT WAS MADE TO REPLACE WITH A 4196 LEAD BUT THAT WAS UNSUCCESSFUL DUE TO CHALLENGES WITH THE PATIENT ANATOMY. AN EPICARDIAL LEAD WILL BE ATTEMPTED IN THE NEAR FUTURE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4196 ASKU

Patients

Seq Age Sex Outcome Treatment
1 75 YR Male Hospitalization| R| O 4035 COMPETITOR IMPLANTABLE PACING LEAD| 4064 COMPETITOR IMPLANTABLE PACING LEAD| D224TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4035 COMPETITOR IMPLANTABLE PACING LEAD| 4064 COMPETITOR IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB