CAPSURE Z NOVUS

MDR report key: 1920797 · Received December 13, 2010

Report

Report Number
2649622-2010-12337
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P850089/S38
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER A REGULAR PACEMAKER CHECK, IT WAS CONFIRMED THE VENTRICULAR LEAD POLARITY HAD BEEN CHANGED BY THE LEAD MONITOR FUNCTION DUE TO LOWERED LEAD IMPEDANCE. THE SUBCLAVIAN FRACTURE WAS CONFIRMED WITH A RADIOSCOPY INSPECTION. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE Z NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5054 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention 5554 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR