FDA Adverse Event
Injury
Summary report: N
CAPSURE Z NOVUS
MDR report key: 1920797
·
Received December 13, 2010
Report
- Report Number
- 2649622-2010-12337
- Event Type
- Injury
- Date Received
- December 13, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P850089/S38
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.
Description of Event or Problem · 1
IT WAS REPORTED THAT AFTER A REGULAR PACEMAKER CHECK, IT WAS CONFIRMED THE VENTRICULAR LEAD POLARITY HAD BEEN CHANGED BY THE LEAD MONITOR FUNCTION DUE TO LOWERED LEAD IMPEDANCE. THE SUBCLAVIAN FRACTURE WAS CONFIRMED WITH A RADIOSCOPY INSPECTION. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE Z NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5054 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | 5554 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR |