FDA Adverse Event Injury Summary report: N

ENTRUST AT

MDR report key: 1920783 · Received December 13, 2010

Report

Report Number
6000144-2010-05652
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S46
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE DEVICE LASTED 71% OF ITS EXPECTED LONGEVITY. BOTH LOADED AND UNLOADED PACING CURRENT DRAIN LEVELS WERE NORMAL FOR THIS DEVICE OVER THE RANGE OF BATTERY VOLTAGE IT OPERATED UNDER DURING ITS SERVICE TIME. THERE IS NO EVIDENCE TO INDICATE A PROBLEM WITH THE BATTERY. WITHOUT KNOWING THE PROGRAMMING HISTORY OF THIS DEVICE THERE IS NO WAY TO DETERMINE WHY IT DID NOT MEET ITS EXPECTED LONGEVITY CALCULATION.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED DUE TO NORMAL BATTERY DEPLETION. THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTRUST AT IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D153ATG ASKU

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| R 4076 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD