CONCERTO CRT-D DR
Report
- Report Number
- 6000144-2010-05651
- Event Type
- Injury
- Date Received
- December 13, 2010
- Date of Event
- June 23, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S031
- Removal / Correction Number
- Z-0126-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DC, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS FOUND THE CAUSE OF THE HIGH CURRENT DRAIN TO BE CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.
THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. THE DEVICE WAS RETURNED WITHOUT INFORMATION REGARDING PRODUCT PERFORMANCE. THE MANUFACTURER'S DEVICE TRACKING SYSTEM INDICATES THAT THE DEVICE WAS REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONCERTO CRT-D DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | NIK | MEDTRONIC MED REL, INC. | C154DWK | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Hospitalization| R | 4193 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD| 4076 IMPLANTABLE PACING LEAD |