SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA
Report
- Report Number
- 2023826-2010-01071
- Event Type
- Malfunction
- Date Received
- November 2, 2010
- Date of Event
- October 5, 2010
- Report Date
- October 5, 2010
- Manufacturer
- STAAR SURGICAL COMPANY
- Product Code
- HQL
- PMA / PMN Number
- P900048
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- NURSE
Narratives
(B)(4): EVAL RESULTS: VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE LENS OPTIC IS TORN. LENS WAS RETURNED IN LENS CASE. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. CONCLUSIONS: AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED IN JUNE 2005 (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MANUFACTURING OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY AND THE EVAL OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).
THE REPORTER STATED THE SURGEON INSERTED AN AQ2015A SILICONE ASPHERIC THREE PIECE LENS. LENS TORE ON INSERTION INTO THE EYE. LENS WAS REMOVED WITH NO PT INJURY. ANOTHER SAME MODEL AND SIZE LENS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA | INTRAOCULAR LENS | HQL | STAAR SURGICAL COMPANY | AQ2015A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARTRIDGE: MODEL - AQ CARTRIDGE-FP, LOT NUMBER UNK| INJECTOR: MODEL MSI - TM, LOT NUMBER UNK |