FDA Adverse Event Malfunction Summary report: N

TALENT CAPTIVIA THORACIC STENT GRAFT SYSTEM

MDR report key: 1920717 · Received December 9, 2010

Report

Report Number
2953200-2010-02502
Event Type
Malfunction
Date Received
December 9, 2010
Date of Event
November 14, 2010
Report Date
November 14, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL, RESULTS: BOVINE THORACIC ARCH AND ANGULATED AORTA. STENT GRAFT IMPLANTED IN PT WITH INADEQUATE ANATOMY. EVAL, CONCLUSION: BOVINE THORACIC ARCH AND ANGULATED AORTA. STENT GRAFT IMPLANTED IN PT WITH INADEQUATE ANATOMY.

Description of Event or Problem · 1

A TALENT CAPTIVIA STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 7 CM LONG CHRONIC DISSECTING TAA. THE PT HAD A BOVINE ARCH THAT WAS EXTREMELY TORTUOUS, AND THERE WAS A 110 DEGREE ANGULATION IN THE TRANSVERSE AORTA BEFORE THE DESCENDING AORTA. ONE DEVICE WAS IMPLANTED IN THE ARCH AND LANDED AS PLANNED AT THE INNOMINATE ARTERY; HOWEVER, DUE TO THE SEVERE AORTIC ANGULATION, THERE WAS A KINK IN THE GRAFT BETWEEN THE FIRST AND SECOND COVERED STENTS ALONG THE LESSER CURVE OF THE AORTA. THE WIRE WAS ALSO HUGGING THE INNER AORTIC CURVE. DURING REMOVAL OF THE DELIVERY SYSTEM, THE PHYSICIAN TRIED TO ADVANCE THE SPINDLE BUT WAS UNABLE TO DUE TO THE INTERACTIONS BETWEEN THE DEVICE AND THE GRAFT AND THE SEVERE ANGLED ANATOMY. HOWEVER, BY PULLING THE TRIGGER AND MOVING THE GRAY HANDLE BACK AND FORTH, THE DELIVERY SYSTEM WAS ABLE TO BE WITHDRAWN WITHOUT ISSUES. THE KINKED GRAFT WAS LEFT UNCORRECTED. NO CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE. THE DELIVERY SYSTEM WAS DISCARDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT CAPTIVIA THORACIC STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00558641

Patients

Seq Age Sex Outcome Treatment
1 60 YR