FDA Adverse Event
Malfunction
Summary report: N
M SERIES
MDR report key: 1920666
·
Received December 6, 2010
Report
- Report Number
- 1220908-2010-03424
- Event Type
- Malfunction
- Date Received
- December 6, 2010
- Report Date
- November 15, 2010
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- K990762
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORP; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY INTEGRATED CIRCUIT ON THE BRIDGE BOARD. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "BRIDGE TEST FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | M SERIES | DEFIBRILLATOR/PACEMAKER | MKJ | ZOLL MEDICAL CORPORATION | MSERIES BI-PHASIC | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |