FDA Adverse Event Malfunction Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1920603 · Received October 28, 2010

Report

Report Number
2023826-2010-01081
Event Type
Malfunction
Date Received
October 28, 2010
Date of Event
September 1, 2010
Report Date
October 5, 2010
Manufacturer
STAAR SURGICAL CO.
Product Code
MTA
PMA / PMN Number
P9800048
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).EVAL RESULTS (OTHER): VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE LENS OPTIC TORN, WITH A PIECE TORN OFF. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. CONCLUSIONS (OTHER): BASED ON THE COMPLAINT HISTORY AND THE EVAL OF THE RETURNED PRODUCT, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, WERE ADDRESSED IN A CAPA OPENED IN (B)(4) 2005 (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MFG OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. (B)(4).

Description of Event or Problem · 1

THE REPORTER STATED THE SURGEON ATTEMPTED TO INSERT AN AQ2015A SILICONE ASPHERIC THREE PIECE LENS, BUT THE LENS WOULDN'T EJECT OUT OF THE CARTRIDGE PROPERLY AND TORE. THERE WAS PT CONTACT, BUT NO INJURY. ANOTHER SAME MODEL LENS WAS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS MTA STAAR SURGICAL CO. AQ2015A NA

Patients

Seq Age Sex Outcome Treatment
1 INJECTOR MODEL MSI-TM -LOT # UNK| CARTRIDGE MODEL AQ CARTRIDGE -FP- LOT # UNK