FDA Adverse Event Malfunction Summary report: N

BD EMERALD SYRINGE

MDR report key: 19205995 · Received April 29, 2024

Report

Report Number
2243072-2024-00576
Event Type
Malfunction
Date Received
April 29, 2024
Date of Event
April 2, 2024
Report Date
June 25, 2024
Manufacturer
BD MEDICAL, MEDICATION DELIVERY SOLUTIONS
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

A QUALITY ENGINEER INSPECTED THE 1 PHYSICAL SAMPLE AND 01 PHOTOGRAPH OF LOT 2061289 REGARDING THE ITEM #303080 SUBMITTED FOR EVALUATION. ANALYSIS OF THE SAMPLES & PHOTOGRAPH SHOWED THAT THE CONTAMINATION IS VISIBLE IN THE SYRINGE. A REVIEW OF OUR RISK MANAGEMENT DOCUMENT WAS PERFORMED AND INDICATES THAT THE POTENTIAL RISK OF THE REPORTED EVENT WAS ASSESSED APPROPRIATELY. PRODUCTION RECORDS WERE REVIEWED, AND THIS BATCH MEETS OUR SPECIFICATION REQUIREMENTS. BASED ON THE RETURN SAMPLES. THE REPORTED ISSUE OF ¿CONTAMINATION¿ WAS CONFIRMED. THE PHOTO OF EMERALD SYRINGE 5ML LOT NO. 2061289 RECEIVED AND EVALUATED. THE IMAGE SHOWING THE CONTAMINATION IS VISIBLE IN THE SYRINGE. QUALITY ENGINEER INVESTIGATE AND DID SIMULATION ON THE RETAINED SAMPLES ON OUR END FOR THE LOT NO. 2061289, ITEM #303080. BUT THERE WAS NOT ANY DEFECT OBSERVED IN RETAINED SAMPLES. THE DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NO. 2061289. ITEM # 303080. THERE WAS NO REJECTED INSPECTIONS OR QUALITY NOTIFICATIONS DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER. ROOT CAUSE : THE LINE OPERATOR AND TECHNICIANS WERE INTERVIEWED FOR THE POSSIBILITY OF DUST FALLING INTO PRODUCT CAVITY. PROBABLE ROOT CAUSE FOR THIS DEFECT WAS FOUND TO BE IMPROPER CLEANING OF SYRINGE LOADERS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD EMERALD SYRINGE HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. THE SYRINGE POCKET HAVE A LOT OF DUST INSIDE AS IT HAS HOLES IN IT AND IT SEEMS ITS THE ONLY SYRINGE SEEMS TO BE HAVING DUST AS THE OTHER SYRINGES ARE IN GOOD CONDITION.

Description of Event or Problem · 0

THE SYRINGE POCKET HAVE A LOT OF DUST INSIDE AS IT HAS HOLES IN IT AND IT SEEMS ITS THE ONLY SYRINGE SEEMS TO BE HAVING DUST AS THE OTHER SYRINGES ARE IN GOOD CONDITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1278412 BD EMERALD SYRINGE INTRAVASCULAR CATHETER FOZ BD MEDICAL, MEDICATION DELIVERY SOLUTIONS 2061289

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown