BD EMERALD SYRINGE
Report
- Report Number
- 2243072-2024-00576
- Event Type
- Malfunction
- Date Received
- April 29, 2024
- Date of Event
- April 2, 2024
- Report Date
- June 25, 2024
- Manufacturer
- BD MEDICAL, MEDICATION DELIVERY SOLUTIONS
- Product Code
- FOZ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
A QUALITY ENGINEER INSPECTED THE 1 PHYSICAL SAMPLE AND 01 PHOTOGRAPH OF LOT 2061289 REGARDING THE ITEM #303080 SUBMITTED FOR EVALUATION. ANALYSIS OF THE SAMPLES & PHOTOGRAPH SHOWED THAT THE CONTAMINATION IS VISIBLE IN THE SYRINGE. A REVIEW OF OUR RISK MANAGEMENT DOCUMENT WAS PERFORMED AND INDICATES THAT THE POTENTIAL RISK OF THE REPORTED EVENT WAS ASSESSED APPROPRIATELY. PRODUCTION RECORDS WERE REVIEWED, AND THIS BATCH MEETS OUR SPECIFICATION REQUIREMENTS. BASED ON THE RETURN SAMPLES. THE REPORTED ISSUE OF ¿CONTAMINATION¿ WAS CONFIRMED. THE PHOTO OF EMERALD SYRINGE 5ML LOT NO. 2061289 RECEIVED AND EVALUATED. THE IMAGE SHOWING THE CONTAMINATION IS VISIBLE IN THE SYRINGE. QUALITY ENGINEER INVESTIGATE AND DID SIMULATION ON THE RETAINED SAMPLES ON OUR END FOR THE LOT NO. 2061289, ITEM #303080. BUT THERE WAS NOT ANY DEFECT OBSERVED IN RETAINED SAMPLES. THE DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NO. 2061289. ITEM # 303080. THERE WAS NO REJECTED INSPECTIONS OR QUALITY NOTIFICATIONS DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER. ROOT CAUSE : THE LINE OPERATOR AND TECHNICIANS WERE INTERVIEWED FOR THE POSSIBILITY OF DUST FALLING INTO PRODUCT CAVITY. PROBABLE ROOT CAUSE FOR THIS DEFECT WAS FOUND TO BE IMPROPER CLEANING OF SYRINGE LOADERS.
IT WAS REPORTED THAT BD EMERALD SYRINGE HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. THE SYRINGE POCKET HAVE A LOT OF DUST INSIDE AS IT HAS HOLES IN IT AND IT SEEMS ITS THE ONLY SYRINGE SEEMS TO BE HAVING DUST AS THE OTHER SYRINGES ARE IN GOOD CONDITION.
THE SYRINGE POCKET HAVE A LOT OF DUST INSIDE AS IT HAS HOLES IN IT AND IT SEEMS ITS THE ONLY SYRINGE SEEMS TO BE HAVING DUST AS THE OTHER SYRINGES ARE IN GOOD CONDITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1278412 | BD EMERALD SYRINGE | INTRAVASCULAR CATHETER | FOZ | BD MEDICAL, MEDICATION DELIVERY SOLUTIONS | 2061289 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |