FDA Adverse Event Malfunction Summary report: N

COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH

MDR report key: 1920411 · Received December 10, 2010

Report

Report Number
6000001-2010-05749
Event Type
Malfunction
Date Received
December 10, 2010
Date of Event
September 24, 2010
Report Date
October 7, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE DEVICE WAS EVALUATED ONSITE AT THE FACILITY BY A BAXTER TECHNICIAN. A VISUAL INSPECTION AND FUNCTIONAL TESTS WERE PERFORMED. THE REPORTED CONDITION OF A BATTERY ISSUE WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE DEPLETED MAIN BATTERIES. A BAXTER REPAIR TECHNICIAN REPLACED THE BATTERIES AND HARNESS TO RESOLVE THIS ISSUE. THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). THIS INFORMATION WAS INCORRECT IN THE INITIAL MEDWATCH. ADDITIONAL EVALUATION CODE THAT WAS NOT INCLUDED IN INITIAL MEDWATCH.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT A COLLEAGUE INFUSION PUMP EXPERIENCED A BATTERY ISSUE. BAXTER'S REVIEW OF THE DEVICE EVENT HISTORY DETERMINED A BATTERY DEPLETED SET INTERRUPTED DELIVERY. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED. THE MASTER SOFTWARE VERSION OF THIS DEVICE WAS 5.80.92, WHICH IS CLASSIFIED AS REMEDIATED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1