6461500 INFUSOR PUMP (1 LABEL)
Report
- Report Number
- 6000001-2010-05735
- Event Type
- Malfunction
- Date Received
- December 10, 2010
- Date of Event
- September 1, 2010
- Report Date
- September 29, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K885377
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE CAUSE WAS AN OUT OF ADJUSTMENT OCCLUSION SWITCH. THE OCCLUSION SWITCH HAS BEEN READJUSTED. ADDITIONAL: A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS NOT PREVIOUSLY SERVICED PRIOR TO THIS EVENT.
(B)(4). THE ALERT AND OCCURENCE DATES WERE INACCURATE IN THE INITIAL MEDWATCH REPORT. THE ACCURATE DATE FOR BOTH THE ALERT DATE AND OCCURRENCE DATE IS (B)(6) 2010.
(B)(4), BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
THE FACILITY REPRESENTATIVE REPORTED A INFUSOR PUMP WITH A CONDITION OF "DROPPED IT AND NOW IT WON'T WORK". IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. DURING BAXTER'S PRODUCT EVALUATION, IT WAS DISCOVERED THAT THERE WAS A FALSE OCCLUSION ALARM. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 6461500 INFUSOR PUMP (1 LABEL) | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |