FDA Adverse Event Injury Summary report: N

RX HERCULINK 14 PLUS STENT SYSTEM

MDR report key: 1920327 · Received December 10, 2010

Report

Report Number
3004742046-2010-00601
Event Type
Injury
Date Received
December 10, 2010
Date of Event
November 17, 2010
Report Date
November 17, 2010
Manufacturer
AV-TEMECULA
Product Code
FGE
PMA / PMN Number
K063481
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE TWO RX HERCULINK STENT SYSTEMS (PART 1008021-12, LOT 0062251 AND PART 1008021-15, LOT UNK) INDICATED ARE BEING FILED UNDER SEPARATE MEDWATCH MFR NUMBERS. (B)(4): VASCULAR USE OF A BILIARY STENT. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS NOT PROVIDED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED PATIENT EFFECT OF DISSECTION IS A KNOWN ADVERSE EVENT LISTED IN THE FAILURE MODE AND EFFECTS ANALYSIS AS A KNOWN HAZARD. ALTHOUGH A CONCLUSIVE CAUSE FOR THE REPORTED PATIENT EFFECT AND THE RELATIONSHIP TO THE DEVICE, IF ANY, CANNOT BE DETERMINED, THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY WITH RESPECT TO MANUFACTURE, OR DESIGN. REPORTEDLY, THE HERCULINK PLUS STENT DELIVERY SYSTEM WAS USED TO TREAT A RENAL ARTERY LESION. IT SHOULD BE NOTED THAT THE HERCULINK PLUS IFU STATES THAT THE RX HERCULINK PLUS BILIARY STENT SYSTEM IS INTENDED FOR PALLIATION OF MALIGNANT STRICTURES IN THE BILIARY TREE. IN THIS CASE, IT IS UNKNOWN IF THE OFF LABEL USE OF THE PRODUCT CONTRIBUTED TO THE REPORTED DISSECTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE PROCEDURE AT THE OSTIUM OF THE RENAL ARTERY, A 5.5X12 HERCULINK PLUS STENT SYSTEM WAS ADVANCED AND DEPLOYED SUCCESSFULLY AT THE TARGET LESION; HOWEVER, A DISSECTION OCCURRED AT THE DISTAL END OF THE STENT. A 6.0X12 HERCULINK PLUS STENT SYSTEM WAS ADVANCED AND AN ATTEMPT WAS MADE TO OVERLAP THE PREVIOUS STENT TO COVER THE DISSECTION. THE BALLOON WAS INFLATED TO 10 ATMOSPHERES (ATM) AND THEN THE BALLOON WAS DEFLATED. UPON DEFLATION OF THE BALLOON THE STENT JUMPED 30MM DISTAL. THE HERCULINK PLUS BALLOON WAS ADVANCED AND WAS ABLE TO RETRIEVE THE STENT TO THE SITE OF THE DISSECTION. THE BALLOON WAS INFLATED TO 14ATM AND THE STENT AGAIN JUMPED DISTAL. A 6.0X15 HERCULINK PLUS STENT SYSTEM WAS ADVANCED TO THE SITE OF THE DISSECTION AND THE BALLOON WAS INFLATED TO 10ATM. THE BALLOON WAS DEFLATED AND THE STENT JUMPED DISTAL. A 8.0X40 NON ABBOTT STENT SYSTEM WAS ADVANCED IN AN EFFORT TO COVER THE DISSECTION, BUT THE STENT SYSTEM COULD NOT CROSS AND WAS REMOVED FROM THE ANATOMY. THE TWO STENTS THAT REMAINED DISTAL TO THE DISSECTION WERE RETRIEVED USING THE HERCULINK PLUS BALLOON AND PLACED AT THE SITE OF THE DISSECTION. THE 6.0X12 STENT REMAINED 3MM FROM THE 5.5X12 STENT AND THE 6.0X15 STENT REMAINED 3MM FROM THE 6.0X12 STENT. THE PHYSICIAN STATED THAT THE DISSECTION CAUSED THE VESSEL TO BECOME AN 8MM VESSEL AND THAT IS THE REASON WHY THE STENTS WERE NOT ADHERING TO THE VESSEL. THE PHYSICIAN ALSO STATED THAT ONCE THE DISSECTION HEALS, THE STENTS WILL ADHERE TO THE VESSEL. THERE WERE NO ADVERSE PATIENT EFFECTS THROUGHOUT OR POST PROCEDURE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RX HERCULINK 14 PLUS STENT SYSTEM BILIARY STENT SYSTEM FGE AV-TEMECULA

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention