FDA Adverse Event Malfunction Summary report: N

FREESTYLE NAVIGATOR

MDR report key: 1920289 · Received December 10, 2010

Report

Report Number
2954323-2010-01669
Event Type
Malfunction
Date Received
December 10, 2010
Date of Event
November 19, 2010
Report Date
March 1, 2011
Product Code
LFR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SENT ONCE INVESTIGATION RESULTS ARE AVAILABLE. REMEDIAL ACTION 2954323-04/14/2009-002-C WAS REPORTED TO THE (B)(4) OFFICE ON 14 APR 2009. NOTE: IT SHOULD BE NOTED THAT ON (B)(6) 2010 CUSTOMER ORIGINALLY COMPLAINED OF CRACKS NEAR THE BATTERY COMPARTMENT OF TRANSMITTER (B)(4). PRODUCT RETURN WAS REQUESTED AND THE PRODUCT WAS NEVER RECEIVED.

Additional Manufacturer Narrative · 1

RETURNED TRANSMITTER WAS VISUALLY INSPECTED AND CONFIRMED TO HAVE CRACKS IN THE THERMISTOR AND BATTERY DOOR LATCH AREAS. LEAK TESTING WAS ALSO PERFORMED. THE RETURNED TRANSMITTER PASSED LEAK TESTING. BASED ON THESE RESULTS, THIS ISSUE IS NOT ASSOCIATED WITH REMEDIAL ACTION 2954323-04/14/2009-002-C.

Description of Event or Problem · 1

A CUSTOMER REPORTED THEY OBSERVED A FRACTURE IN THEIR FREESTYLE NAVIGATOR TRANSMITTER. IT HAS BEEN DETERMINED THAT SOME FREESTYLE NAVIGATOR TRANSMITTERS COULD POTENTIALLY EXHIBIT A CRACK/FRACTURE ON THE PLASTIC HOUSING NEAR THE BATTERY COMPARTMENT, WHICH COULD POTENTIALLY ALLOW MOISTURE TO COME IN CONTACT WITH THE TRANSMITTER'S INTERNAL ELECTRONICS, POTENTIALLY CAUSING THE TRANSMITTER AND THE RECEIVER TO LOSE CONNECTION INTERRUPTING THE CONTINUOUS GLUCOSE RESULTS. ALTHOUGH UNLIKELY, MOISTURE ENTERING THE TRANSMITTER HAS THE POTENTIAL TO GENERATE INACCURATE RESULTS ONLY WITH THE CONTINUOUS GLUCOSE READINGS. THE BUILT-IN FREESTYLE GLUCOSE METER IS NOT IMPACTED BY THIS ISSUE AND THE USER CAN CONTINUE TO USE THE BUILT-IN METER. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.