CYPHER SIROLIMUS-ELUTING CORONARY STENT
Report
- Report Number
- 9616099-2010-00937
- Event Type
- Malfunction
- Date Received
- December 10, 2010
- Report Date
- November 22, 2010
- Manufacturer
- CORDIS DE MEXICO
- Product Code
- NIQ
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE LOCATION OF THE STENT FRACTURE WAS UNKNOWN AND NO TREATMENT WAS PROVIDED. POPMA ET AL CLASSIFICATION WAS IV (COMPLETE SEPARATION WITH DISPLACEMENT). ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THIS ADVERSE EVENT IN WHICH ASSOCIATED MANUFACTURER REPORT NUMBERS ARE 9616099-2010-00936 AND 9616099-2010-00937. (B)(4) CYPHER PRODUCT (CJS) IS NOT DISTRIBUTED IN THE US; HOWEVER, IT IS SIMILAR TO US DISTRIBUTED CYPHER PRODUCT (CXS).
ADDITIONAL INFORMATION WAS RECEIVED ON (B)(4) 2010. ONLY 1 OF THE 2 DEVICES WAS FRACTURED. A STENT FRACTURE, WITH SEPARATION AND DISPLACEMENT, AND 28% RESTENOSIS WAS CONFIRMED FOLLOWING IMPLANTATION OF TWO CYPHER BX STENTS. THIS COMPLAINT IS FROM THE AMERICAN HEART JOURNAL 2010; 160:775.E1-775.E9 ARTICLE ENTITLED 'SERIAL ANGIOGRAPHIC FINDINGS AND PROGNOSIS OF STENT FRACTURE SITE WITHOUT EARLY RESTENOSIS AFTER SIROLIMUS-ELUTING STENT IMPLANTATION.' THIS CASE IS 16TH OF 29 EVENTS. THE PATIENT WAS A (B)(6) MALE. THE INDICATION FOR THE INDEX PROCEDURE WAS STABLE CORONARY ARTERY DISEASE. THE TARGET LESION WAS LEFT ANTERIOR DESCENDING (LAD) ARTERY, BUT THE DETAILS ARE NOT INDICATED. THE LESION WAS NEITHER AN IN-STENT RESTENOSIS NOR A CHRONIC TOTAL OCCLUSION. THE DIAMETER OF THE VESSEL AND THE LESION LENGTH WAS UNKNOWN. PRE-PROCEDURE STENOSIS WAS UNKNOWN. NO FURTHER INFORMATION AVAILABLE. BEFORE THE EVENT OCCURRED, TWO CYPHER BX (SIZE AND LOT UNKNOWN) STENTS WERE IMPLANTED WITH OVERLAP. THE TOTAL STENT LENGTH WAS 32MM AND THE STENT DIAMETER WAS 2.5MM. THE DATE OF THE PROCEDURE AND THE DETAILS OF THE STENT IMPLANTATION WERE UNKNOWN. THE FIRST FOLLOW-UP ANGIOGRAM WAS PERFORMED SIX TO NINE MONTHS AFTER THE IMPLANT AND STENT FRACTURE WAS OBSERVED. THERE WAS NO BINARY RESTENOSIS, BUT IT WAS REPORTED THAT THERE WAS 28% STENOSIS. THE LOCATION OF THE STENT FRACTURE WAS UNKNOWN AND NO TREATMENT WAS PROVIDED. POPMA ET AL CLASSIFICATION WAS IV (COMPLETE SEPARATION WITH DISPLACEMENT).THE PRODUCTS REMAIN IMPLANTED IN THE PATIENT AND ARE THUS NOT AVAILABLE FOR EVALUATION. A REVIEW OF THE MANUFACTURING RECORDS COULD NOT BE CONDUCTED WITHOUT A LOT NUMBER. ADDITIONAL INFORMATION RECEIVED INDICATED THAT ONLY ONE OF THE TWO STENTS WAS FRACTURED. REOCCLUSION/RESTENOSIS IS A KNOWN POTENTIAL ADVERSE EVENT ASSOCIATED WITH CORONARY ARTERY STENTING AND IS OFTEN ASSOCIATED WITH THE PROGRESSION OF CARDIOVASCULAR DISEASE. STENT DAMAGE/DEFORMATION AND RESTENOSIS/REOCLUSSION ARE KNOWN ADVERSE EVENTS INDICATED IN THE IFU. WHILE NOT OBSERVED IN THE PIVOTAL CLINICAL TRIALS THAT SUPPORTED THE CYPHER STENT PMA, STENT FRACTURES ARE UNCOMMON EVENTS BUT HAVE BEEN OBSERVED IN LONG STENTED SEGMENTS INCLUDING THOSE IN WHICH OVERLAPPING STENTS HAVE BEEN USED. THEY HAVE BEEN OBSERVED IN CORONARY SEGMENTS THAT UNDERGO SIGNIFICANT MOTION, PARTICULARLY IN AREAS WITH SEVERE ANGULATION, TORTUOSITY AND CALCIFICATION. IN THE CYPHER STENT, THEY HAVE BEEN REPORTED MOST OFTEN IN CERTAIN LESION SUBGROUPS IN WHICH SAFETY AND EFFECTIVENESS HAVE NOT BEEN ESTABLISHED. RECENTLY, SES FRACTURE HAS BEEN RECOGNIZED AS A NEW POTENTIAL MECHANISM OF RESTENOSIS. POSSIBLE DISPOSING FACTORS OF STENT FRACTURE ARE OVERLAPPING STENTS, CARDIAC MOTION AND VESSEL ANGULATIONS/TORTUOSITY FOR THE NATIVE CORONARY ARTERY. BASED ON THE LIMITED PROCEDURAL INFORMATION AND WITHOUT FILMS, IT IS NOT POSSIBLE TO DRAW A DEFINITIVE CONCLUSION REGARDING THE REPORTED EVENTS. HOWEVER, VESSEL/LESION CHARACTERISTICS AND PROCEDURAL FACTORS ARE LIKELY CONTRIBUTING FACTORS. THERE IS NO INDICATION THAT THE EVENTS ARE RELATED TO THE DEVICE MANUFACTURING PROCESS. THEREFORE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. THIS IS ONE OF TWO PRODUCTS INVOLVED WITH THIS ADVERSE EVENT IN WHICH ASSOCIATED MANUFACTURER REPORT NUMBERS ARE 9616099-2010-00936 AND 9616099-2010-00937. (B)(4) CYPHER PRODUCT (CJS) IS NOT DISTRIBUTED IN THE US; HOWEVER, IT IS SIMILAR TO US DISTRIBUTED CYPHER PRODUCT (CXS).
A STENT FRACTURE, WITH SEPARATION AND DISPLACEMENT, AND 28% RESTENOSIS WAS CONFIRMED FOLLOWING IMPLANTATION OF TWO CYPHER BX STENTS. THIS COMPLAINT IS FROM THE AMERICAN HEART JOURNAL 2010; 160:775.E1-775.E9 ARTICLE ENTITLED "SERIAL ANGIOGRAPHIC FINDINGS AND PROGNOSIS OF STENT FRACTURE SITE WITHOUT EARLY RESTENOSIS AFTER SIROLIMUS-ELUTING STENT IMPLANTATION." THIS CASE IS 16TH OF 29 EVENTS. THE PATIENT WAS A (B)(6) MALE. THE INDICATION FOR THE INDEX PROCEDURE WAS STABLE CORONARY ARTERY DISEASE. THE TARGET LESION WAS LEFT ANTERIOR DESCENDING (LAD) ARTERY, BUT THE DETAILS ARE NOT INDICATED. THE LESION WAS NEITHER AN IN-STENT RESTENOSIS NOR A CHRONIC TOTAL OCCLUSION. THE DIAMETER OF THE VESSEL AND THE LESION LENGTH WAS UNKNOWN. PRE-PROCEDURE STENOSIS WAS UNKNOWN. NO FURTHER INFORMATION AVAILABLE. BEFORE THE EVENT OCCURRED, TWO CYPHER BX (SIZE AND LOT UNKNOWN) STENTS WERE IMPLANTED WITH OVERLAP. THE TOTAL STENT LENGTH WAS 32MM AND THE STENT DIAMETER WAS 2.5MM. THE DATE OF THE PROCEDURE AND THE DETAILS OF THE STENT IMPLANTATION WERE UNKNOWN. THE FIRST FOLLOW-UP ANGIOGRAM WAS PERFORMED SIX TO NINE MONTHS AFTER THE IMPLANT AND STENT FRACTURE WAS OBSERVED. THERE WAS NO BINARY RESTENOSIS, BUT IT WAS REPORTED THAT THERE WAS 28% STENOSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CYPHER SIROLIMUS-ELUTING CORONARY STENT | DRUG-ELUTING STENT (NIQ) | NIQ | CORDIS DE MEXICO | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |