INRATIO
Report
- Report Number
- 2027969-2010-02098
- Event Type
- Malfunction
- Date Received
- December 2, 2010
- Date of Event
- August 9, 2010
- Report Date
- December 2, 2010
- Manufacturer
- ALERE SAN DIEGO, INC
- Product Code
- GJS
- PMA / PMN Number
- 021923
- Removal / Correction Number
- NI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NURSE
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: INRATIO: 1.6, REFERENCE: 2.2, MEAN: 1.90, CONFIDENCE LIMITS: 1.3-2.7; 3.9, 3.0, 3.45, 2.0-5.0. ANALYSIS OF CUSTOMER'S DATA REVEALED THAT BOTH PTS' INRATIO AND REFERENCE TEST RESULT COMPARISONS MEET ACCURACY CRITERIA. CUSTOMER'S RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. NO PRODUCT IS EXPECTED TO BE RETURNED. NO FURTHER INVESTIGATION REQUIRED. AS REVIEWED ON 12/02/2010, THIRTY-TWO DISCREPANT RESULT COMPLAINTS WERE REPORTED FOR LOT #232889 YIELDING A COMPLAINT RATE OF 0.017%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (>0.07%), NO FURTHER ACTION IS REQUIRED AT THIS TIME. THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.
CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: PT 1: DATE: (B)(6) 2010, INRATIO: 1.6, LAB: 2.2; PT 2: (B)(6) 2010, 3.9, 3.0.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC | 100139 | 232889 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |