FDA Adverse Event Malfunction Summary report: N

BD INSYTE 18GA X 1.16IN

MDR report key: 19193737 · Received April 26, 2024

Report

Report Number
9610048-2024-00041
Event Type
Malfunction
Date Received
April 26, 2024
Date of Event
March 29, 2024
Report Date
August 28, 2024
Manufacturer
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
Product Code
FOZ
UDI-DI
00382903883172
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
PHARMACIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 38831714 AND LOT NUMBER 3194195 & 3241011. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS SAMPLES WERE UNAVAILABLE FOR THIS EVENT, A THOROUGH SAMPLE ANALYSIS COULD NOT BE COMPLETED. ALTHOUGH AN EXACT CAUSE COULD NOT BE IDENTIFIED FOR THIS INCIDENT, A CORRECTIVE AND PREVENTIVE ACTION PLAN HAS BEEN INITIATED TO FURTHER INVESTIGATE THIS ISSUE AND PREVENT ANY REOCCURRENCE. OUR QUALITY TEAM WILL CONTINUE TO CLOSELY MONITOR THE MANUFACTURING PROCESS FOR SIGNS OF THIS POTENTIAL DEFECT AND ANY EMERGING TRENDS.

Additional Manufacturer Narrative · 0

THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. OTHER POTENTIAL LOT NUMBERS INCLUDE 3194195 WITH EXPIRATION DATE 2028-07-31, AND 3241011 WITH EXPIRATION DATE 09/30/2028. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED. OTHER POTENTIAL LOT NUMBER DEVICE MANUFACTURE DATES INCLUDE 2023-08-09 AND 10/04/2023.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INSYTE 18GA X 1.16IN NEEDLE WENT THROUGH THE CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER TRANSLATED FROM SPANISH TO ENGLISH: "WE HAVE RECEIVED NOTIFICATION OF ADVERSE INCIDENTS WITH THE BD INSYTE INTRAVENOUS CATHETER OF 24, 18, 20 GAUGE. IN PARTICULAR, WE HAVE BEEN INFORMED THAT THE IV CATHETER NO. 24 LOT 3062305 FV 02/2028 AT THE MOMENT OF UNCLOGGING THE CATHETER, THE CATHETER WAS CHECKED, AND THE MANDREL WAS OBSERVED TO BE PERFORATED BY THE NEEDLE. THE ICU SERVICE REPORTS AN INCREASE IN CASES OF MECHANICAL PHLEBITIS." ADDITIONAL INFORMATION RECEIVED APR 09 2024 WHAT IS THE QUANTITY AFFECTED? RT/ THE NOVELTY OF THE DEVICES HAS BEEN PRESENTED IN THE CATHETER NO. 18,22,20 AND 24 IN THE AREAS OF UCI EMERGENCY AND PEDIATRICS. HOWEVER, 2 CASES HAVE BEEN FORMALLY REPORTED WITH CATHETER NO. 24 AND 2 CASES WITH CATHETER NO. 18. WHAT IS THE LOT NUMBER OF THE PRODUCT? R/ CATHETER 18 LOT: 3194195 EXPIRY DATE 30/07/2028 - 3241011 EXPIRY DATE 33/08/2028 WAS THE REPORTED INCIDENT DETECTED BEFORE, DURING OR AFTER USE ON THE PATIENT? R/ BEFORE USE ON PATIENT AND IN USE ON PATIENT, GENERATING PHLEBITIS. HAS THERE BEEN ANY HARM TO THE PATIENT/HEALTHCARE PROFESSIONAL? R/ PHLEBITIS HAS OCCURRED IN THE VEINS OF A MULTI-PUNCTURED PATIENT. - WAS THERE A NEED FOR MEDICAL AND/OR SURGICAL INTERVENTION DUE TO WHAT HAPPENED (IMAGING TESTS, SURGERY, MEDICATION ADMINISTRATION, ETC.)? (DETAIL) FOR THE MOMENT NO - HAS THERE BEEN EXPOSURE OF BLOOD/CHEMOTHERAPY TO MUCOUS MEMBRANES (EYES, NOSE AND MOUTH) OR SKIN? IF YES, INDICATE WHETHER THE EXPOSURE WAS FROM THE PATIENT OR THE PRACTITIONER AND WHAT MEASURES WERE TAKEN. (DETAIL) FOR THE TIME BEING NO - WHAT MEDICATION WAS BEING ADMINISTERED? R/ MEDICATION WAS NOT ADMINISTERED AS THERE WAS EVIDENCE OF DAMAGE AT THE TIME OF CHANNELING THE PATIENT. - IS THE SAMPLE RELATED TO THE INCIDENT AVAILABLE FOR ANALYSIS? IF YES, REPORT THE QUANTITY. R/ THE CATHETERS HAVE BEEN DISCARDED. - COULD YOU PLEASE SEND PHOTO SAMPLES OF THE SPECIMEN? THE FOLLOWING PHOTOS ARE SENT AS EVIDENCE: - O DATE OF INCIDENT: MARCH 29, MARCH 30 AND APRIL 2 OF THE CURRENT YEAR:

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
871569 BD INSYTE 18GA X 1.16IN PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. UNKNOWN 00382903883172

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown