FDA Adverse Event Malfunction Summary report: N

TPUC3

MDR report key: 19192430 · Received April 26, 2024

Report

Report Number
1823260-2024-01289
Event Type
Malfunction
Date Received
April 26, 2024
Date of Event
March 25, 2024
Report Date
June 25, 2024
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JGQ
UDI-DI
04015630919536
PMA / PMN Number
K141925
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE SERIAL NUMBER OF THE COBAS 6000 C501 MODULE WAS NOT PROVIDED. THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative · 0

BASED ON THE INFORMATION PROVIDED, THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Description of Event or Problem · 0

THE INITIAL REPORTER RECEIVED QUESTIONABLE HIGH TPUC3 TOTAL PROTEIN URINE/CSF GEN.3 RESULTS FROM ONE URINE PATIENT SAMPLE TESTED ON THE COBAS 6000 C501 MODULE. THE INITIAL TPUC3 RESULT OF THE UNDILUTED PATIENT SAMPLE FROM THE MODULE WAS 152.6 MG/DL. THE REPEAT RESULT WITH THE PATIENT SAMPLE AT 1:3 DILUTION FROM THE MODULE WAS 1193.4 MG/DL. THE REPEAT RESULT OF THE UNDILUTED PATIENT SAMPLE FROM THE MODULE WAS 132.6 MG/DL. THE REPEAT RESULT USING AN IMMUNOTURBIDIMETRY LABORATORY METHOD WAS 15.6 MG/L. THE REPORTER SENT THE PATIENT SAMPLE TO AN OUTSOURCED LABORATORY. THE REPEAT RESULT FROM A COMPETITOR ANALYZER USING AN IMMUNOTURBIDITY LABORATORY METHOD WAS 5.5 MG/L. THE PATIENT SAMPLE WAS ALSO SENT TO THE INVESTIGATION LABORATORY. THE REPEAT RESULT USING A BENZONTIUM CHLORIDE METHOD WAS 155.2 MG/DL. THE INVESTIGATION LABORATORY SUSPECTS THE PRESENCE OF AN INTERFERENT IN THE PATIENT SAMPLE. ADDITIONAL TPUC3 RESULTS OF 22.4 MG/DL AND 397.8 MG/DL WERE PROVIDED. IT IS NOT CLEAR HOW THESE RESULTS RELATE TO THE EVENT. CLARIFICATION HAS BEEN REQUESTED RELATED TO THESE RESULTS BUT HAS NOT BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2122375 TPUC3 TURBIDIMETRIC, TOTAL PROTEIN JGQ ROCHE DIAGNOSTICS 75158401 04015630919536

Patients

Seq Age Sex Outcome Treatment
1 4 YR Male