FDA Adverse Event Malfunction Summary report: N

EASYPOINT

MDR report key: 19191230 · Received April 26, 2024

Report

Report Number
19191230
Event Type
Malfunction
Date Received
April 26, 2024
Date of Event
April 3, 2024
Report Date
April 4, 2024
Manufacturer
RETRACTABLE TECHNOLOGIES, INC.
Product Code
FMI
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

IN THE PAST ALMOST 2 YEARS, OUR FACILITY HAS HAD A TREMENDOUS NUMBER OF ISSUES WITH EASYPOINT NEEDLES. THEY ARE UNSAFE FOR MEDICAL PROFESSIONALS AND PATIENTS. ISSUES THAT WE HAVE HAD: NEEDLE NOT RETRACTING, BLOOD SPRAY WITH RETRACTION OF NEEDLE, NEEDLE POPPING OFF SYRINGE WITH ATTEMPTED RETRACTION WITH SYRINGE POPPING OUT, DULL NEEDLES, FAILING TO PUNCTURE SKIN, CAUSING PAIN, NEEDLES BENDING WHEN DRAWING UP MEDS, ABNORMAL CURVATURE OF TIP OF NEEDLE, SYRINGE/NEEDLE COMPLETELY FALLING APART UPON RETRACTION, PROTECTIVE CAP FALLING OFF WHEN REMOVED FROM PACKAGE, NEEDLES HAVING CONNECTION ISSUES, NOT TIGHTENING, LEAKING AT THE HUB. WE HAVE GONE THROUGH THE PROPER CHANNELS, HAD RETRACTABLE TECHNOLOGY, INC (RTI) EDUCATORS OUT TO MEET WITH STAFF, SENT BACK DEFECTIVE NEEDLES, AND STILL NO RESOLUTION. OUR SUPPLY CHAIN DEPARTMENT HAS OVER 55 REPORTS, AND THIS DOESN'T ACCOUNT FOR THE EVENTS THAT OCCUR THAT ARE NOT REPORTED.

Description of Event or Problem · 0

IN THE PAST ALMOST 2 YEARS, OUR FACILITY HAS HAD A TREMENDOUS NUMBER OF ISSUES WITH EASYPOINT NEEDLES. THEY ARE UNSAFE FOR MEDICAL PROFESSIONALS AND PATIENTS. ISSUES THAT WE HAVE HAD: - NEEDLE NOT RETRACTING - BLOOD SPRAY WITH RETRACTION OF NEEDLE - NEEDLE POPPING OFF SYRINGE WITH ATTEMPTED RETRACTION WITH SYRINGE POPPING OUT - DULL NEEDLES, FAILING TO PUNCTURE SKIN, CAUSING PAIN - NEEDLES BENDING WHEN DRAWING UP MEDS - ABNORMAL CURVATURE OF TIP OF NEEDLE - SYRINGE/NEEDLE COMPLETELY FALLING APART UPON RETRACTION - PROTECTIVE CAP FALLING OFF WHEN REMOVED FROM PACKAGE - NEEDLES HAVING CONNECTION ISSUES, NOT TIGHTENING - LEAKING AT THE HUB WE HAVE GONE THROUGH THE PROPER CHANNELS, HAD RETRACTABLE TECHNOLOGY, INC (RTI) EDUCATORS OUT TO MEET WITH STAFF, SENT BACK DEFECTIVE NEEDLES, AND STILL NO RESOLUTION. OUR SUPPLY CHAIN DEPARTMENT HAS OVER 55 REPORTS, AND THIS DOESN'T ACCOUNT FOR THE EVENTS THAT OCCUR THAT ARE NOT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2123309 EASYPOINT NEEDLE, HYPODERMIC, SINGLE LUMEN FMI RETRACTABLE TECHNOLOGIES, INC. 82011 K201207
435404 EASYPOINT NEEDLE, HYPODERMIC, SINGLE LUMEN FMI RETRACTABLE TECHNOLOGIES, INC. 82011 K201207

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown