FDA Adverse Event Injury Summary report: N

AU5431-02 CLINICAL CHEMISTRY ANALYZER

MDR report key: 1918984 · Received December 10, 2010

Report

Report Number
2050012-2010-01558
Event Type
Injury
Date Received
December 10, 2010
Date of Event
November 5, 2010
Report Date
November 10, 2010
Manufacturer
BECKMAN COULTER MISHIMA K.K.
Product Code
JJE
PMA / PMN Number
K011720
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

QC WAS WITHIN RANGE PRIOR TO THE EVENT BUT WAS OUT OF RANGE AFTER THE EVENT. A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE AND PERFORMED TROUBLESHOOTING. FSE INDICATED THAT THE PROBLEM WAS DUE TO A BAD SAMPLE VALVE FOR THE OUTER WHEEL AND REPLACED THE SAMPLE VALVE. THE SAMPLE VALVE WAS ORIGINAL TO THE INSTRUMENT AND WAS ON THE SYSTEM FOR AT LEAST 9 YEARS.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING ERRONEOUSLY HIGH IRON RESULTS FOR 24 PATIENTS GENERATED BY THE AU5431-02 CLINICAL CHEMISTRY ANALYZER. THE RESULTS WERE REPORTED OUTSIDE THE LABORATORY AND CORRECTED REPORTS WERE ISSUED AFTER REPEAT TESTING WAS PERFORMED. IT IS UNKNOWN IF PATIENT TREATMENT WAS AFFECTED IN THIS EVENT. THE RISK OF SERIOUS INJURY WILL BE ASSUMED UPON RECUR. A SEPARATE REPORT 2050012-2010-01660 HAS BEEN SUBMITTED FOR THE MALFUNCTION PORTION OF THIS EVENT AND THE REPORTS SHOWN BELOW HAVE BEEN SUBMITTED FOR THE OTHER PATIENTS ASSOCIATED WITH THIS EVENT. 2050012-2010-01540; 2050012-2010-01556; 2050012-2010-01557; 2050012-2010-01559; 2050012-2010-01560; 2050012-2010-01561; 2050012-2010-01562; 2050012-2010-01563; 2050012-2010-01564; 2050012-2010-01565; 2050012-2010-01566; 2050012-2010-01567; 2050012-2010-01568; 2050012-2010-01569; 2050012-2010-01570; 2050012-2010-01571; 2050012-2010-01572; 2050012-2010-01573; 2050012-2010-01574; 2050012-2010-01575; 2050012-2010-01576; 2050012-2010-01577; 2050012-2010-01578.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AU5431-02 CLINICAL CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER MISHIMA K.K. NA NA

Patients

Seq Age Sex Outcome Treatment
1 Other