FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 19188958 · Received April 26, 2024

Report

Report Number
9610877-2024-53081
Event Type
Malfunction
Date Received
April 26, 2024
Date of Event
April 18, 2024
Report Date
April 26, 2024
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
PMA / PMN Number
K172156
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE WITH DRIVE PCB SHADOW IN IMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CCD MODULE WITH DRIVE PCB. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE LCB DISTAL COVER GLASS CRACKED, THE INSERTION FLEXIBLE TUBE CRUSHED, AND THE REMOTE CONTROL BUTTONS WORN OUT; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(SHADOW IN IMAGE).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2126445 PENTAX VIDEO NASO PHARYNGO LARYNGOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE VNL8-J10

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown