FDA Adverse Event Malfunction Summary report: N

RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS

MDR report key: 19188395 · Received April 25, 2024

Report

Report Number
9610773-2024-01107
Event Type
Malfunction
Date Received
April 25, 2024
Report Date
September 25, 2024
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HIH
UDI-DI
04042761023658
PMA / PMN Number
K931994
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE EVALUATION IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PMA/510(K) NUMBER: K931994/K931995.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE CUSTOMER'S REPORTABLE MALFUNCTION WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. IT HAS BEEN OVER 1 YEAR SINCE THE SUBJECT DEVICE WAS MANUFACTURED. BASED ON THE RESULTS OF THE INVESTIGATION, THE REPORTED ISSUE WAS LIKELY CAUSED BY THERMAL MECHANICAL OVERLOAD, IMPROPER HANDLING, MECHANICAL IMPACT LIKE FALL, SHOCK OR SIMILAR STRESS. HOWEVER, A ROOT CAUSE COULD NOT BE DETERMINED. THE EVENT CAN BE DETECTED & PREVENTED BY FOLLOWING THE INSTRUCTIONS FOR USE (IFU) WHICH STATE: CHAPTER 4.1 ¿INSPECTION AND TESTING¿ OF THE INSTRUCTIONS FOR USE (IFU) REGARDING THE PROPER REPROCESSING OF THE AFFECTED DEVICE. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS REPORTED, THE SHEATH TIP WAS CHIPPED. THE ISSUE OCCURRED, DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1039586 RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS HYSTEROSCOPE (AND ACCESSORIES) HIH OLYMPUS WINTER & IBE GMBH A42011A 23205 04042761023658

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown