FDA Adverse Event Malfunction Summary report: N

GALILEO® EXTRACTION TOOL, CAPTURING ROD

MDR report key: 19179675 · Received April 25, 2024

Report

Report Number
1220246-2024-02350
Event Type
Malfunction
Date Received
April 25, 2024
Date of Event
April 1, 2024
Report Date
September 4, 2024
Manufacturer
ARTHREX, INC.
Product Code
HSB
UDI-DI
00848665030891
PMA / PMN Number
K021008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

THE COMPLAINT ALLEGATION WAS CONFIRMED. ONE UNPACKAGED 0857-100 BATCH 221019 WAS RECEIVED FOR INVESTIGATION. THE DEVICE ARRIVED STUCK INSIDE THE PART NUMBER 1099-0090 BATCH 212625. FUNCTIONAL TESTING WAS NOT PERFORMED DUE TO THE DEVICE ARRIVED STUCK INSIDE THE PART NUMBER 1099-0090. UPON VISUAL INSPECTION, DAMAGE TO THE SHAFT AND THE THREADED TIP WAS OBSERVED. IT WAS ALSO NOTED THAT THE DEVICE WAS BENT. THE MOST LIKELY CAUSE(S) OF THE REPORTED FAILURE ARE USER ERROR, MISALIGNED INSERTION, OR EXCESSIVE FORCE DURING USE.

Additional Manufacturer Narrative · 0

THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD NOT BE DETERMINED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION. IF THE DEVICE BECOMES AVAILABLE FOR EVALUATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

ON (B)(6) 2024, DURING DEVICE EVALUATION OF THE RETURNED DEVICE, IT WAS FOUND THAT THE 0857-100 WAS STUCK INSIDE A 1099-090 GALILEO LAG SCREW, RIGHT, Ø10.5MMX90MM.

Description of Event or Problem · 0

ON 04/01/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-06529172 THAT AN 0857-100 GALILEO EXTRACTION TOOL WAS DAMAGED DURING THE CASE. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1468101 GALILEO® EXTRACTION TOOL, CAPTURING ROD INTRAMED FIXATN ROD AND ACCESS HSB ARTHREX, INC. GALILEO® EXTRACTION TOOL, CAPTURING ROD UNK 00848665030891

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown