FDA Adverse Event Malfunction Summary report: N

MICROCLAVE® CLEAR NEUTRAL CONNECTOR

MDR report key: 19177077 · Received April 24, 2024

Report

Report Number
9617594-2024-00497
Event Type
Malfunction
Date Received
April 24, 2024
Date of Event
March 1, 2024
Report Date
June 18, 2024
Manufacturer
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
Product Code
FPA
UDI-DI
00887709037151
PMA / PMN Number
K100434
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL PHONE # FROM E1 IS (B)(6). THE DEVICE IS AVAILABLE FOR EVALUATION- IT HAS NOT BEEN RECEIVED.

Additional Manufacturer Narrative · 0

THE COMPLAINT OF LEAKS ON ITEM MC100 CANNOT BE CONFIRMED. SINCE NO PRODUCT SAMPLES, PICTURES, OR VIDEOS WERE RECEIVED FOR INVESTIGATION. WITHOUT THE RETURN OF THE USED SAMPLE, A COMPREHENSIVE FAILURE INVESTIGATION CANNOT BE PERFORMED AND PROBABLE CAUSE CANNOT BE DETERMINED. THE DEVICE HISTORY REPORT (DHR) FOR LOT#: 5960803 WAS REVIEWED AND NO NON-CONFORMITIES WERE FOUND THAT WOULD HAVE LED TO THE REPORTED CONDITION ON THE COMPLAINT.

Description of Event or Problem · 0

THE EVENT OCCURRED ON AN UNKNOWN DATE INVOLVING A MICROCLAVE® CLEAR NEUTRAL CONNECTOR WHERE THE CUSTOMER REPORTED THAT THE PATIENT'S PERIPHERALLY INSERTED CENTRAL CATHETER (PICC) LINE CLOTTED AND HAD TO BE REMOVED BECAUSE THE HEPARIN INFUSION WAS LEAKING. THE NURSE NOTICED THE LINE WAS LEAKING SHORTLY AFTER SETTING UP THE LINES. THE NURSE ADDED CLAVES TO DETERMINE IF THAT WOULD REMEDY THE PROBLEM; HOWEVER, THE LEAKING CONTINUED. THE NURSE THEN CHANGED ALL OF THE CLAVES, BUT THAT DID NOT SOLVE THE ISSUE. EVENTUALLY, THE NURSE HAD TO CHANGE THE ENTIRE LINE SETUP. THERE WAS UNKNOWN PATIENT INVOLVEMENT AND NO PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
502279 MICROCLAVE® CLEAR NEUTRAL CONNECTOR SET, ADMINISTRATION, INTRAVASCULAR FPA ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 5960803 00887709037151

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown HEPARIN, MFR UNK.| PERIPHERALLY INSERTED CENTRAL CATHETER, MFR UNK.