FDA Adverse Event Malfunction Summary report: N

SYRINGE 5ML SALINE FILL CHINA SP

MDR report key: 19173844 · Received April 24, 2024

Report

Report Number
1911916-2024-00281
Event Type
Malfunction
Date Received
April 24, 2024
Date of Event
March 30, 2024
Report Date
May 17, 2024
Manufacturer
BECTON DICKINSON
Product Code
FMF
UDI-DI
30382903065944
PMA / PMN Number
K161552
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. H3 OTHER TEXT : SEE H10 MANUFACTURE NARRATIVE.

Additional Manufacturer Narrative · 0

(B)(4) FOLLOW UP. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 306594 AND LOT NUMBER 3090847. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. THE PRODUCT IS DESIGNED TO PUSH THE PLUNGER ROD DOWN TO EXPEL THE SOLUTION. IT IS NOT FOR PULLING THE PLUNGER ROD BACK. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME.

Description of Event or Problem · 0

AFTER USE, THE NURSE PULLS BACK WITH THE HAND AND THE PIN IS SEPARATED FROM THE SKIN PLUG.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED. AFTER USE, THE NURSE PULLS BACK WITH THE HAND AND THE PIN IS SEPARATED FROM THE SKIN PLUG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1084727 SYRINGE 5ML SALINE FILL CHINA SP SYRINGE, PISTON FMF BECTON DICKINSON 3090847 30382903065944

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown