FDA Adverse Event Malfunction Summary report: N

BD BBL¿ SENSI-DISC¿ PIPERACILLIN/TAZOBACTAM - 100/10 ¿G

MDR report key: 19173661 · Received April 24, 2024

Report

Report Number
2647876-2024-00067
Event Type
Malfunction
Date Received
April 24, 2024
Date of Event
March 28, 2024
Report Date
June 19, 2024
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
JTN
UDI-DI
00382902316916
PMA / PMN Number
K940359
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: A COMPLAINT INVESTIGATION DUE TO BREAKPOINTS LISTED IN THE PACKAGE INSERT (PI) FOR PIPERACILLIN/TAZOBACTAM CATALOG 231691 (BATCH NUMBER NOT RECEIVED) ARE ALIGNED WITH CLSI M100:ED30 INSTEAD OF THE M100:ED33. THE INVESTIGATION REQUIRED TO EVALUATE PRODUCT INSERT AND INFORMATION SENT BY CUSTOMER. INFORMATION SENT BY CUSTOMER (PHOTO FROM PRODUCT INSERT SECTION AND CLSI INFORMATION) WERE EVALUATED. IT WAS IDENTIFIED THAT PRODUCT INSERT REQUIRED REVISION. COMPLAINT IS CONFIRMED. AS A CORRECTIVE ACTION PRODUCT INSERT WAS SUBMITTED FOR REVISION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED.

Additional Manufacturer Narrative · 0

H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PRODUCT INSERT FOR BD BBL¿ SENSI-DISC¿ PIPERACILLIN/TAZOBACTAM - 100/10 G CONTAINED INCORRECT BREAKPOINT INFORMATION FOR PIPERACILLIN/TAZOBACTAM. THERE WAS NO REPORT OF PATIENT IMPACT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PRODUCT INSERT FOR BD BBL¿ SENSI-DISC¿ PIPERACILLIN/TAZOBACTAM - 100/10 G CONTAINED INCORRECT BREAKPOINT INFORMATION FOR PIPERACILLIN/TAZOBACTAM. THERE WAS NO REPORT OF PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1273521 BD BBL¿ SENSI-DISC¿ PIPERACILLIN/TAZOBACTAM - 100/10 ¿G SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL JTN BECTON DICKINSON CARIBE LTD. UNKNOWN 00382902316916

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown