FDA Adverse Event Other Summary report: N

DA VINCI SI SURGICAL SYSTEM

MDR report key: 1916752 · Received December 7, 2010

Report

Report Number
2955842-2010-00534
Event Type
Other
Date Received
December 7, 2010
Date of Event
November 8, 2010
Report Date
November 8, 2010
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K081137
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONDUCTED BY THE FIELD SERVICE ENGINEER CONCLUDED THAT SYSTEM ERROR CODE 23008 WAS ASSOCIATED WITH THE LEFT MASTER TOOL MANIPULATOR (MTML) GIMBAL. THE MTM GIMBAL IS A SUBSECTION OF THE COMPLETE MTM ARM, CONSISTING OF THE LINKS ASSOCIATED WITH THE FIVE AXES OF MOTION CLOSEST TO THE SURGEON'S HAND. THESE AXES MEASURE THE ORIENTATION OF THE MTM HANDLE AND THE MTM GRIP ANGLE. THE GIMBAL IS SEPARATED FROM THE REST OF THE MTM BY REMOVING THE PLATFORM LINK FROM THE FOREARM LINK. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED MTM GIMBAL. THE SYSTEM ALARM (SYSTEM GENERATED FAULT CODE) FUNCTIONED AS DESIGNED AND THERE WAS NO INJURY TO THE PATIENT. SYSTEM ERROR CODE 23008 APPEARS WHEN THE DA VINCI S SAFETY SYSTEM DETERMINES THAT THE ANGULAR POSITION OF ONE OR MORE ROBOTIC JOINTS ON THE SPECIFIED MANIPULATOR, AS MEASURED BY THAT JOINT'S PRIMARY CONTROL SENSOR (ENCODER) AND THE SECONDARY SENSOR (POTENTIOMETER), WERE OUT OF SPECIFIED TOLERANCE FOR AGREEMENT. UPON DETERMINING THIS CONDITION, THE SAFETY SYSTEM PUTS DA VINCI IN A NONRECOVERABLE SAFE STATE. THE MTML GIMBAL WAS RETURNED TO ISI FOR FAILURE ANALYSIS INVESTIGATION. ENGINEERING CONFIRMED THE REPORTED COMPLAINT AND REPLACED THE MOTOR TO REPAIR THE AFFECTED MTML. AS OF DECEMBER 6, 2009, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S HYSTERECTOMY PROCEDURE, THE SITE EXPERIENCED SYSTEM ERROR 23008. WITH THE ASSISTANCE OF AN ISI TECHNICAL SUPPORT ENGINEER (TSE), THE SITE EXERCISED THE AFFECTED ARM AXIS TO RECOVER FROM THE SYSTEM ERROR AND RESTARTED THE SYSTEM; HOWEVER, THE FAULT IMMEDIATELY RETURNED. AT THIS TIME, THE SURGEON MADE THE DECISION TO CONVERT TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO COMPLETE THE PLANNED SURGICAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI SI SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL,INC. IS3000 A6.0P5

Patients

Seq Age Sex Outcome Treatment
1 Other DA VINCI SI SYSTEM INSTRUMENT, ACCESSORIES & ESU