DA VINCI SI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2010-00534
- Event Type
- Other
- Date Received
- December 7, 2010
- Date of Event
- November 8, 2010
- Report Date
- November 8, 2010
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K081137
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION CONDUCTED BY THE FIELD SERVICE ENGINEER CONCLUDED THAT SYSTEM ERROR CODE 23008 WAS ASSOCIATED WITH THE LEFT MASTER TOOL MANIPULATOR (MTML) GIMBAL. THE MTM GIMBAL IS A SUBSECTION OF THE COMPLETE MTM ARM, CONSISTING OF THE LINKS ASSOCIATED WITH THE FIVE AXES OF MOTION CLOSEST TO THE SURGEON'S HAND. THESE AXES MEASURE THE ORIENTATION OF THE MTM HANDLE AND THE MTM GRIP ANGLE. THE GIMBAL IS SEPARATED FROM THE REST OF THE MTM BY REMOVING THE PLATFORM LINK FROM THE FOREARM LINK. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED MTM GIMBAL. THE SYSTEM ALARM (SYSTEM GENERATED FAULT CODE) FUNCTIONED AS DESIGNED AND THERE WAS NO INJURY TO THE PATIENT. SYSTEM ERROR CODE 23008 APPEARS WHEN THE DA VINCI S SAFETY SYSTEM DETERMINES THAT THE ANGULAR POSITION OF ONE OR MORE ROBOTIC JOINTS ON THE SPECIFIED MANIPULATOR, AS MEASURED BY THAT JOINT'S PRIMARY CONTROL SENSOR (ENCODER) AND THE SECONDARY SENSOR (POTENTIOMETER), WERE OUT OF SPECIFIED TOLERANCE FOR AGREEMENT. UPON DETERMINING THIS CONDITION, THE SAFETY SYSTEM PUTS DA VINCI IN A NONRECOVERABLE SAFE STATE. THE MTML GIMBAL WAS RETURNED TO ISI FOR FAILURE ANALYSIS INVESTIGATION. ENGINEERING CONFIRMED THE REPORTED COMPLAINT AND REPLACED THE MOTOR TO REPAIR THE AFFECTED MTML. AS OF DECEMBER 6, 2009, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT DURING A DA VINCI S HYSTERECTOMY PROCEDURE, THE SITE EXPERIENCED SYSTEM ERROR 23008. WITH THE ASSISTANCE OF AN ISI TECHNICAL SUPPORT ENGINEER (TSE), THE SITE EXERCISED THE AFFECTED ARM AXIS TO RECOVER FROM THE SYSTEM ERROR AND RESTARTED THE SYSTEM; HOWEVER, THE FAULT IMMEDIATELY RETURNED. AT THIS TIME, THE SURGEON MADE THE DECISION TO CONVERT TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO COMPLETE THE PLANNED SURGICAL PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI SI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS3000 A6.0P5 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | DA VINCI SI SYSTEM INSTRUMENT, ACCESSORIES & ESU |