FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEMS
MDR report key: 1916706
·
Received December 7, 2010
Report
- Report Number
- 2050012-2010-01541
- Event Type
- Malfunction
- Date Received
- December 7, 2010
- Date of Event
- November 10, 2010
- Report Date
- November 10, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
QC PRIOR TO THE EVENT WAS WITHIN LAB-ESTABLISHED RANGES. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED: THE FSE REPLACED THE CARBON BRIDGE AND VERIFIED PERFORMANCE OF THE SYSTEM. A CLEAR ROOT CAUSE FOR THIS EVENT IS UNKNOWN.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING FALSE HIGH SODIUM (NA) AND CHLORIDE (CL) RESULTS GENERATED BY THE UNICEL DXC 600 PRO SYNCHRON CLINICAL SYSTEMS FOR TWO PATIENTS. THE ORIGINAL SAMPLES WERE RETESTED AND OBTAINED RESULTS WERE LOWER AND WERE REPORTED OUT OF THE LAB. THE FALSE HIGH RESULTS WERE NOT REPORTED OUT OF THE LAB. PATIENT TREATMENT WAS NOT AFFECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600 PRO SYNCHRON® CLINICAL SYSTEMS | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600 PRO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |