FDA Adverse Event Malfunction Summary report: N

I-STAT EC8+ CARTRIDGE

MDR report key: 19166439 · Received April 23, 2024

Report

Report Number
2245578-2024-00081
Event Type
Malfunction
Date Received
April 23, 2024
Date of Event
April 16, 2024
Report Date
June 4, 2024
Manufacturer
ABBOTT POINT OF CARE
Product Code
CHL
PMA / PMN Number
K940918
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT (B)(4) APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 28-MAY-2024. A REVIEW OF THE DEVICE HISTORY RECORDS (DHRS) CONFIRMED THE CARTRIDGE LOTS MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AN, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN DETERMINED FOR EG7+ CARTRIDGE LOT N23305, CHEM8+ CARTRIDGE LOT H23287 OR EC8+ CARTRIDGE LOT K24009.

Description of Event or Problem · 0

ON (B)(6) 2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EC8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT POTASSIUM RESULTS ON A 60 YEAR OLD MALE PATIENT WITH ABDOMINAL PAIN AND BLOOD IN URINE. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECT TEST POTASSIUM SAMPLE I-STAT ON (B)(6) 2024 08:30 08:40 >9.0 A . I-STAT ON (B)(6) 2024 08:30 08:50 >9.0 A . I-STAT ON (B)(6) 2024 08:30 09:00 7.4 A . I-STAT ON (B)(6) 2024 09:10 09:15 4.4 B , GEM/3500 ON (B)(6) 2024 08:30 09:30 4.2 A . GEM/3500 ON (B)(6) 2024 09:10 09:35 3.9 B . AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1189590 I-STAT EC8+ CARTRIDGE EC8+ CARTRIDGE CHL ABBOTT POINT OF CARE NA K24009

Patients

Seq Age Sex Outcome Treatment
1 NA Male