FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 19163964 · Received April 23, 2024

Report

Report Number
9617229-2024-06984
Event Type
Injury
Date Received
April 23, 2024
Date of Event
February 21, 2023
Report Date
August 23, 2024
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
05060191606752
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF CAPSULAR CONTRACTURE AND RUPTURE WAS RECEIVED ON (B)(6) 2024, WITH LOT NUMBER 2012569. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS ARE: CAPSULAR CONTRACTURE: UNABLE TO OBSERVE AS IT IS NOT RELATED TO THE DEVICE. RUPTURE: NOT OBSERVED OPENINGS DURING THE ANALYSIS. AS PER THE INVESTIGATION PROCEDURES CREASES, WEAR ABRASION, PARTICLES AND DEFORMATION WERE COMPLETED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT.

Additional Manufacturer Narrative · 0

FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Description of Event or Problem · 0

PATIENT REPORTED LEFT SIDE SUSPECTED DEVICE RUPTURE. HEALTHCARE PROFESSIONAL LATER REPORTED CAPSULAR CONTRACTURE BAKER GRADE III. THE DEVICE HAS BEEN EXPLANTED AND REPLACED WITH ANOTHER MANUFACTURER'S DEVICE.

Description of Event or Problem · 0

PATIENT REPORTED LEFT SIDE SUSPECTED DEVICE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1328577 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) NI 05060191606752

Patients

Seq Age Sex Outcome Treatment
1 44 YR Female Required Intervention