FDA Adverse Event Injury Summary report: N

REZOOM

MDR report key: 1915761 · Received December 6, 2010

Report

Report Number
2648035-2010-00236
Event Type
Injury
Date Received
December 6, 2010
Date of Event
November 17, 2010
Report Date
November 8, 2010
Manufacturer
ABBOTT MEDICAL OPTICS
Product Code
MFK
PMA / PMN Number
P960028/SO11
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INTRAOCULAR LENS (IOL) ASSOCIATED WITH THIS REPORT WAS SENT BY THE PHYSICIAN TO AN INDEPENDENT LABORATORY FOR ANALYSIS, RESULTS ARE UNKNOWN AT THIS TIME. IN FOLLOW-UP WITH THE SURGEON, IT WAS LEARNED THE PATIENT HAD A MYOPIC SHIFT OF 2.0 DIOPTERS AND WAS REPORTING GHOSTING OF IMAGES. THE ANTERIOR CHAMBER DEPTH OF THE PATIENT'S EYE WAS REDUCED COMPARED TO THE OTHER IMPLANTED EYE. ALL INFORMATION AVAILABLE IS INCLUDED IN THIS REPORT. IOL NOT RETURNED TO THE MANUFACTURER.

Additional Manufacturer Narrative · 1

THE INTRAOCULAR LENS WAS NOT RECEIVED BY THE MANUFACTURER FOR ANALYSIS. SEVERAL ATTEMPTS WERE MADE TO OBTAIN THE TEST RESULTS FROM THE SURGEON WITHOUT SUCCESS. THE MANUFACTURING DOCUMENTATION SHOWS THAT THE PRODUCTION ORDER WAS MANUFACTURED ACCORDING TO SPECIFICATIONS. A REVIEW OF THE POST MARKET SURVEILLANCE DATABASE SHOWS NO OTHER REPORTS RECEIVED FOR LENSES MANUFACTURED IN THIS LOT. A ROOT CAUSE FOR THIS EVENT COULD NOT BE DETERMINED. ALL INFORMATION AVAILABLE IS CONTAINED IN THIS REPORT.

Description of Event or Problem · 1

INITIAL REPORT RECEIVED FROM THE SURGEON ON (B)(6) 2010 WAS AN INQUIRY ASKING IF AN INTRAOCULAR LENS(IOL) WAS IMPLANTED ANTERIOR SIDE TO THE POSTERIOR COULD IT CAUSE A MYOPIC SHIFT OF -2.0 DIOPTERS. IT WAS LATER LEARNED THAT THE IOL WAS EXPLANTED BY THE SURGEON ON (B)(6) 2010 WITHOUT COMPLICATION AND SENT TO AN INDEPENDENT LABORATORY FOR ANALYSIS. PATIENT'S VISUAL ACUITY AT THE TIME OF EXPLANT WAS 6/9 WITH MINIMAL GHOSTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REZOOM MULTIFOCAL INTRAOCULAR LENS MFK ABBOTT MEDICAL OPTICS NXG1

Patients

Seq Age Sex Outcome Treatment
1 59 YR Required Intervention