HARMONIC ACE 23 CM
Report
- Report Number
- 1527736-2008-00626
- Event Type
- Malfunction
- Date Received
- February 2, 2008
- Date of Event
- November 25, 2007
- Report Date
- December 4, 2007
- Manufacturer
- ETHICON ENDO-SURGERY, INC. (CINCINNATI)
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT WHEN ATTEMPTING TO ACTIVATE THE DEVICE ON A GENERATOR GAVE AN ERROR CODE 5. VISUAL EXAMINATION FOUND AN AREA OF THE BLADE THAT WAS DISCOLORED DUE TO HEAT GENERATED FROM CONTACT BETWEEN THE BLADE AND TISSUE PAD. FURTHER ANALYSIS FOUND A CRACK PROPAGATING FROM THE HEAT-AFFECTED AREA OF THE BLADE. THIS FAILURE IS CAUSED DUE TO ACTIVATION OF THE DEVICE WHILE CLAMPED WITHOUT TISSUE BEING PRESENT. FOR THIS REASON THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. THIS CAN RESULT IN DAMAGE TO THE INSTRUMENT." COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LOBECTOMY AN ERROR OCCURRED SOON. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 23 CM | LFL | ETHICON ENDO-SURGERY, INC. (CINCINNATI) | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HANDPIECE| GENERATOR |