ODYSSEY
Report
- Report Number
- 2032139-2010-00005
- Event Type
- Malfunction
- Date Received
- November 4, 2010
- Date of Event
- October 5, 2010
- Report Date
- November 4, 2010
- Manufacturer
- PERMEDICS, INC.
- Product Code
- MUJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
AN URGENT MEDICAL DEVICE CORRECTION LETTER WILL BE DISTRIBUTED TO ALL AFFECTED CUSTOMERS BY (B)(6), 2010, WITH A DESCRIPTION OF THE ANOMALY AND USER CORRECTIVE ACTION STEPS. THE LETTER WILL ALSO BE DISTRIBUTED TO THE PERMEDICS SALES ORGANIZATIONS, INFORMING THEM OF THE ISSUE. A MANDATORY UPGRADE WILL BE PROVIDED BY (B)(6), 2011 AT NO CHARGE. IN THE MEANTIME, THE FOLLOWING RECOMMENDATIONS ARE PROVIDED: IF, PRIOR TO THE SOFTWARE CORRECTION, THE USER CHANGED THE PT'S TREATMENT START DATE ON ANY FULLY-COMPUTED IMRT PLAN, IMAT PLAN OR PLAN CONTAINING A "REGION DVH" PRESCRIPTION, THE USER SHOULD FORCE A FULL RE-COMPUTATION OF THOSE PT PLANS PRIOR TO FIXING AND EXPORTING THE PLAN. THE FOLLOWING PROCEDURE CAN BE USED TO FORCE A FULL RE-COMPUTATION OF THE PLAN: OPEN THE PT'S PLAN WORKSHEET. CHANGE THE ALGORITHM TYPE OF THE PLAN. CHANGE THE ALGORITHM TYPE BACK TO ITS ORIGINAL VALUE. CLOSE THE PLAN WORKSHEET. RE-COMPUTE THE PLAN.
AN ANOMALY WAS FOUND IN THE ODYSSEY SOFTWARE: IF THE TREATMENT START DATE IS CHANGED FOR A PT'S FULLY COMPUTED IMRT PLAN, IMAT PLAN OR A PLAN CONTAINING A "REGION DVH" PRESCRIPTION, THEN ODYSSEY AUTOMATICALLY RE-COMPUTES THE TREATMENT SCHEDULE. HOWEVER, THIS AUTOMATIC RE-COMPUTATION ALSO CHANGES THE MONITOR UNIT VALUES IN THE SCHEDULE, SO THEY ARE NO LONGER CORRECT. IN THESE CASES, IT IS NECESSARY TO MANUALLY PERFORM A FULL RE-COMPUTATION OF THE PLAN TO CALCULATE THE CORRECT MONITOR UNIT VALUES FOR THE TREATMENT SCHEDULE; HOWEVER, ODYSSEY DOES NOT PROVIDE INDICATION TO THE USER OF THIS ANOMALY, AND ALLOWS EXPORT OF THE PLAN WITH THESE INCORRECT MONITOR UNIT VALUES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ODYSSEY | RADIATION TREATMENT PLANNING SYSTEM | MUJ | PERMEDICS, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |