6 SHOOTER SAEED MULTI-BAND LIGATOR
Report
- Report Number
- 1037905-2010-00553
- Event Type
- Malfunction
- Date Received
- November 3, 2010
- Date of Event
- October 5, 2010
- Report Date
- October 5, 2010
- Manufacturer
- COOK ENDOSCOPY
- Product Code
- MND
- PMA / PMN Number
- K944220
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- NURSE
Narratives
EVAL: WE COULD NOT CONFIRM THE REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED TO COOK ENDOSCOPY FOR EVAL. AFTER A REVIEW OF THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED, WE CAN REPORT NO DISCREPANCIES OR ANOMALIES WERE OBSERVED. WE WERE UNABLE TO CONDUCT A SAMPLE TEST FROM THIS LOT BECAUSE NONE OF THE PRODUCT FROM THIS LOT REMAINED IN FINISHED GOODS INVENTORY. A REVIEW OF THE TWELVE MONTH COMPLAINT HISTORY FOR THIS PRODUCT FAMILY WAS CONDUCTED. BASED ON THIS REVIEW, THE LIKELIHOOD OF THIS TYPE OF REPORT IS RARE. CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVAL. THE INSTRUCTIONS FOR USE DIRECT THE USER TO ATTACH THE TRIGGER CORD TO THE HOOK ON THE END OF THE LOADING CATHETER, LEAVING APPROX 2CM OF TRIGGER CORD BETWEEN THE KNOT AND THE HOOK. IF THE KNOT AND THE HOOK ARE PLACED CLOSER THAN 2CM, THIS CAN CREATE RESISTANCE WHEN WITHDRAWING THE LOADING CATHETER THROUGH THE LIGATOR HANDLE AND/OR ACCESSORY CHANNEL OF THE ENDOSCOPE. IF ADDITIONAL PRESSURE IS APPLIED TO THE LOADING CATHETER WHEN RESISTANCE IS ENCOUNTERED, THIS COULD CONTRIBUTE TO SEPARATION OF THE BLUE HOOK FROM THE LOADING CATHETER. PRIOR TO DISTRIBUTION, ALL 6 SHOOTER SAEED MULTI-BAND LIGATORS ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MFG REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. NO FURTHER ACTION WARRANTED AT THIS TIME BECAUSE THE REPORT WAS UNABLE TO BE VERIFIED. THE LIKELIHOOD OF THIS TYPE OF OCCURRENCE IS RARE. CUSTOMER QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.
IN PREPARATION FOR A BAND LIGATION PROCEDURE, THE PHYSICIAN SELECTED A COOK ENDOSCOPY 6 SHOOTER SAEED MULTI-BAND LIGATOR. THE BLUE HOOK SEPARATED FROM ONE END OF THE LOADING CATHETER WHEN THE USER WAS ATTEMPTING TO LOAD THE LIGATOR ON THE ENDOSCOPE (BEFORE CONTACT WITH PT). THE LOADING PROCESS WAS ABLE TO BE COMPLETED BY USING THE OTHER BLUE HOOK ON THE OPPOSITE END OF THE LOADING CATHETER. A SECTION OF THE DEVICE WAS NOT INSIDE THE PT'S BODY. THE PT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 6 SHOOTER SAEED MULTI-BAND LIGATOR | MND, LIGATOR HEMORRHOIDAL | MND | COOK ENDOSCOPY | W2851394 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | OLYMPUS GIF-H180 ENDOSCOPE |