POWERED 60 ECHELON +, 440MM SHAFT
Report
- Report Number
- 3005075853-2024-03123
- Event Type
- Malfunction
- Date Received
- April 22, 2024
- Date of Event
- March 22, 2024
- Report Date
- April 22, 2024
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- UDI-DI
- 10705036014614
- PMA / PMN Number
- K110385
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). DATE SENT: 4/22/2024. D4: BATCH # 714C37. INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO ETHICON FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE DETERMINED THAT ONE PLEE60A DEVICE WAS RETURNED WITH NO APPARENT DAMAGE AND WITH NO RELOAD PRESENT. THE DEVICE WAS RETURNED WITH A NON-WORKING FIRING MECHANISM. NO FURTHER FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. IN ORDER TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS, THE DEVICE WAS DISASSEMBLED. UPON DISASSEMBLING, THE PCB BOARD WAS NOTED TO BE NON-FUNCTIONAL. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE PCB BOARD TO BE DAMAGED. DEVICE HISTORY REVIEW A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE BATCH NUMBER 714C37, AND NO NON-CONFORMANCES WERE IDENTIFIED.
IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE WITHOUT MOTOR SOUND ON THE 1TH FIRING. THE SURGEON REMOVED THE BATTERY AND ROTATED FOR 180 DEGREE AND REINSERTED THE BATTERY. THE DEVICE STILL COULD NOT FIRE AND WITHOUT MOTOR SOUND. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2241776 | POWERED 60 ECHELON +, 440MM SHAFT | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | A9ED5X | 10705036014614 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |