FDA Adverse Event Malfunction Summary report: N

LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM VG

MDR report key: 1915279 · Received November 2, 2010

Report

Report Number
2024601-2010-00855
Event Type
Malfunction
Date Received
November 2, 2010
Date of Event
September 2, 2010
Report Date
October 6, 2010
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MEDWATCH SENT TO FDA ON: (B)(4) 2010. VISUAL EXAM OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ALLERGAN HAS RECEIVED THE PRODUCT; HOWEVER, THE ANALYSIS HAS NOT BEEN COMPLETED AT THIS TIME. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."

Description of Event or Problem · 1

SURGEON REPORTED, "PT DIDN'T GET RESTRICTION OF THE BAND. DURING X-RAY, IT WAS NOTICED THAT THE TUBING WAS COMPLETELY BROKEN. THE PT WAS RE-OPERATED AND GOT A NEW BAND."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM VG LTI ALLERGAN NA NI

Patients

Seq Age Sex Outcome Treatment
1 60 YR