FDA Adverse Event Malfunction Summary report: N

CPT HIP SYSTEM TAPER STEM INSERTER

MDR report key: 1915277 · Received November 2, 2010

Report

Report Number
1822565-2010-01069
Event Type
Malfunction
Date Received
November 2, 2010
Report Date
May 6, 2010
Manufacturer
ZIMMER, INC.
Product Code
LXH
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INFO WAS RECEIVED FROM A FOREIGN SOURCE WHO IS NOT REQUIRED TO COMPLETE FORM 3500A. EVAL SUMMARY: THIS TYPE OF THREAD DAMAGE MIGHT HAVE BEEN CAUSED DUE TO ONE OR COMBINATION OF THE FOLLOWING FACTORS: IF THE DEVICE IS NOT THREADED PROPERLY WITH THE STEM (I.E. CROSS THREADING OR PARTIAL THREAD ENGAGEMENT). OFF AXIS IMPACTION. DAMAGE CAUSED DURING REPEATED CLEANING AND STEAM STERILIZATION. OFF-LABEL USE (E.G. DEVICE IS USED TO EXTRACT THE STEM, HEAVY IMPACTION FORCE, ETC). BALL PLUNGER MIGHT WENT MISSING DURING CLEANING/AUTOCLAVING. HOWEVER, THIS IS NOT AN EXHAUSTIVE LIST. NO DEFINITIVE CAUSE ANALYSIS CAN BE PERFORMED WITH CERTAINTY. EVAL: DEVICE HISTORY RECORDS INDICATE ALL COMPONENTS WERE MFG AND INSPECTED TO SPECIFICATION. AS RETURNED, THE DISTAL THREADS ARE FRACTURED AND THERE ARE IMPACTION MARKS ON THE KNOB. THE BALL PLUNGER IS ALSO MISSING. THE TENTATIVE FIELD AGE OF THE DEVICE IS APPROXIMATELY 5 YRS 3 MONTHS.

Description of Event or Problem · 1

IT IS REPORTED THAT THE DISTAL THREADS OF THE INSERTER HAVE FRACTURED OFF AND THE BALL PLUNGER IS MISSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CPT HIP SYSTEM TAPER STEM INSERTER HIP INSTRUMENT LXH ZIMMER, INC. 60249783

Patients

Seq Age Sex Outcome Treatment
1