- 1423500-2010-06578
- Event Type
- Malfunction
- Date Received
- December 6, 2010
- Date of Event
- November 13, 2010
- Report Date
- November 13, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ENGINEERING QUALITY REVIEW WAS COMPLETED FOR THIS REPORT OF A SYSTEM ERROR 2240 (AIR IN SET). THE REPORTED CONDITION WAS NOT CONFIRMED DUE TO LACK OF SAMPLE. BASED ON THE INFORMATION OBTAINED DURING BAXTER'S INVESTIGATION, THE ROOT CAUSE OF THE ALARM WAS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE PATIENT DISCONNECTED THE PATIENT LINE FROM THE TRANSFER SET. THE HP STATED THAT THE ALARM OCCURRED BECAUSE SHE DISCONNECTED. THE LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. THE BATCH REVIEW WAS NOT PERFORMED BECAUSE THE LOT NUMBER WAS UNKNOWN. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). DURING INVESTIGATION BY BAXTER, THIS INCIDENT WAS DETERMINED TO BE CAUSED BY USE/USER ERROR. THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION; THEREFORE, A BATCH REVIEW AND SAMPLE EVALUATION WILL NOT BE CONDUCTED. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
THE HOME PATIENT (HP) CONTACTED GLOBAL TECHNICAL SERVICES REGARDING A SYSTEM ERROR (SE) 2240 ALARM WHICH OCCURRED ON THE HOMECHOICE (HC) DURING DRAIN 4 OF 5. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) INSTRUCTED THE HP TO RECYCLE POWER TO CLEAR ALARMS. THE TSR FURTHER EXPLAINED SE 2240 INDICATES A LARGE AMOUNT OF AIR HAS ENTERED SETUP AND ADVISED THE HP TO PRESS STOP BEFORE DISCONNECTING TO AVOID THE ALARM. THE HP CONFIRMED SHE WOULD CALL HER NURSE REGARDING THE AIR DETECTION ALARM AND RECONNECTING. ON (B)(6) 2010 PRODUCT SURVEILLANCE SPOKE WITH THE HP WHO STATED THAT THE ALARM OCCURRED BECAUSE SHE DISCONNECTED. THE HP CONFIRMED SHE WAS AWARE OF THE PROPER PROCEDURES ON HOW TO PREVENT THE ALARM FROM OCCURRING. NO ADVERSE EVENT RESULTED FROM THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTEGRATED APD SET W/CASSETTE3-PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |