FDA Adverse Event Malfunction Summary report: N

LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM (UNK SIZE)

MDR report key: 1915201 · Received November 2, 2010

Report

Report Number
2024601-2010-00891
Event Type
Malfunction
Date Received
November 2, 2010
Date of Event
August 23, 2010
Report Date
September 16, 2010
Manufacturer
ALLERGAN
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TAPER II. (B)(4). VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS NOTED THE ACCESS PORT HAD AN UNIDENTIFIED OPENING WITH LEAKAGE. THE PORT TUBING CONTAINED A LINEAR OPENING WITH LEAKAGE AND STRIATIONS CONSISTENT WITH SURGICAL DAMAGE. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."

Description of Event or Problem · 1

HEALTH PROFESSIONAL REPORTED THAT THERE WAS AN EXPLANTED DEVICE IN THEIR POSSESSION THAT NEEDED TO BE RETURNED. NO ADDITIONAL INFO WAS PROVIDED. FURTHER FOLLOW-UP WITH THE HEALTH PROFESSIONAL NOTED "PORT EXPLANT DUE TO LEAK."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM (UNK SIZE) LTI ALLERGAN NA NI

Patients

Seq Age Sex Outcome Treatment
1 41 YR