FDA Adverse Event
Malfunction
Summary report: N
LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM (UNK SIZE)
MDR report key: 1915201
·
Received November 2, 2010
Report
- Report Number
- 2024601-2010-00891
- Event Type
- Malfunction
- Date Received
- November 2, 2010
- Date of Event
- August 23, 2010
- Report Date
- September 16, 2010
- Manufacturer
- ALLERGAN
- Product Code
- LTI
- PMA / PMN Number
- P000008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TAPER II. (B)(4). VISUAL EXAMINATION OF THE RETURNED DEVICE DETERMINED THE ACCESS PORT TUBING CONNECTOR TO BE A TAPER II. ANALYSIS NOTED THE ACCESS PORT HAD AN UNIDENTIFIED OPENING WITH LEAKAGE. THE PORT TUBING CONTAINED A LINEAR OPENING WITH LEAKAGE AND STRIATIONS CONSISTENT WITH SURGICAL DAMAGE. DEVICE LABELING ADDRESSES THE POSSIBLE OUTCOME OF LEAKAGE AS FOLLOWS: "DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTOR TUBING."
Description of Event or Problem · 1
HEALTH PROFESSIONAL REPORTED THAT THERE WAS AN EXPLANTED DEVICE IN THEIR POSSESSION THAT NEEDED TO BE RETURNED. NO ADDITIONAL INFO WAS PROVIDED. FURTHER FOLLOW-UP WITH THE HEALTH PROFESSIONAL NOTED "PORT EXPLANT DUE TO LEAK."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LAP-BAND ADJUSTABLE GASTRIC BANDING SYSTEM (UNK SIZE) | LTI | ALLERGAN | NA | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |