FDA Adverse Event Injury Summary report: N

PROWLER SELECT MICROCATHETERS

MDR report key: 1915177 · Received December 6, 2010

Report

Report Number
1058196-2010-00357
Event Type
Injury
Date Received
December 6, 2010
Date of Event
August 12, 2010
Report Date
November 9, 2010
Manufacturer
CORDIS NEUROVASCULAR, INC.
Product Code
KRA
PMA / PMN Number
K021591
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INFORMATION WAS RECEIVED VIA THE (B)(4) ENTERPRISE PMS THAT THE DAY AFTER AN UNEVENTFUL ENTERPRISE VRD ASSISTED COIL EMBOLIZATION THE PATIENT HAD AN ASYMPTOMATIC CEREBRAL INFARCTION BASED ON IMAGING RESULTS. THE PATIENT WAS ASYMPTOMATIC AND NO TREATMENT WAS PERFORMED. THE PHYSICIAN COMMENTED THAT IT WAS NOT KNOWN IF THE EVENT WAS CAUSED BY THE CATHETER MANIPULATION OR THE ENTERPRISE VRD DEVICE. THE PATIENT WAS REPORTED TO BE CURRENTLY IN STABLE CONDITION. THE INDEX PROCEDURE WAS FOR A COIL EMBOLIZATION OF A CEREBRAL ANEURYSM ASSISTED BY IMPLANTATION OF AN ENTERPRISE VRD STENT. NO TARGET LESION INFORMATION IS AVAILABLE. OTHER IDENTIFIED DEVICES USED DURING THE PROCEDURE INCLUDE A NONCORDIS (CHIKAI) LAUNCHER GUIDING CATHETER, PROWLER SELECT PLUS MICROCATHETER, AND NONCORDIS ECHELON MICROCATHETER WERE USED DURING THE PROCEDURE. THERE IS NO FURTHER PROCEDURAL INFORMATION AVAILABLE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY REPORTED PROBLEMS. THERE WERE NO REPORTED PRODUCT ISSUES, PROCEDURAL COMPLICATIONS OR ADVERSE EVENTS DURING THE PROCEDURE. THE PATIENT HAD NO NEUROLOGICAL DEFICIT UPON LEAVING THE ANGIOGRAPHY SUITE. THE DAY AFTER THE PROCEDURE, THE PATIENT WAS FOUND TO HAVE TWO SPOTS OF PUNCTUATE HIGH INTENSITY AREAS (HIA) IN THE LEFT OCCIPITAL LOBE. IT WAS REPORTED AS AN ASYMPTOMATIC CEREBRAL INFARCTION. THE ENTERPRISE VRD REMAINS IMPLANTED AND THE PROWLER MICROCATHETER IS NOT AVAILABLE FOR ANALYSIS. (B)(4): THE LOT NUMBER OF THE PROWLER SELECT PLUS MICROCATHETER IS NOT KNOWN; THEREFORE A DEVICE HISTORY RECORD REVIEW CANNOT BE COMPLETED. CEREBRAL INFARCTION (STROKE) IS A KNOWN POTENTIAL ADVERSE EVENT ASSOCIATED WITH THIS TYPE OF PROCEDURE AS OUTLINED IN THE INSTRUCTIONS FOR USE. BASED ON THE REPORTED INFORMATION IT IS NOT POSSIBLE TO DRAW A CONCLUSION REGARDING THE CLINICAL RELATIONSHIP BETWEEN THE DEVICES AND THE EVENT. PATIENT, PROCEDURAL, AND PHARMACOLOGICAL FACTORS MAY HAVE CONTRIBUTED TO THE EVENT. THERE IS NO INDICATION OF A RELATIONSHIP TO ANY DEVICE DESIGN, PERFORMANCE, OR MANUFACTURING ISSUES. THEREFORE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. THIS IS ONE OF TWO PRODUCTS USED DURING THE SAME PROCEDURE. PLEASE REFERENCE MFR. REPORT # 1058196-2010-00357 AND # 1058196-2010-00358.

Additional Manufacturer Narrative · 1

INFORMATION WAS RECEIVED VIA (B)(6) THAT THE DAY AFTER AN UNEVENTFUL ENTERPRISE VRD ASSISTED COIL EMBOLIZATION THE PATIENT HAD AN ASYMPTOMATIC CEREBRAL INFARCTION BASED ON IMAGING RESULTS. THE PATIENT WAS ASYMPTOMATIC AND NO TREATMENT WAS PERFORMED. THE PHYSICIAN COMMENTED THAT IT WAS NOT KNOWN IF THE EVENT WAS CAUSED BY THE CATHETER MANIPULATION OR THE ENTERPRISE VRD DEVICE. THE PATIENT WAS REPORTED TO BE CURRENTLY IN STABLE CONDITION. THE INDEX PROCEDURE WAS FOR A COIL EMBOLIZATION OF A CEREBRAL ANEURYSM ASSISTED BY IMPLANTATION OF AN ENTERPRISE VRD STENT. NO TARGET LESION INFORMATION IS AVAILABLE. OTHER IDENTIFIED DEVICES USED DURING THE PROCEDURE INCLUDE A NONCORDIS (CHIKAI) LAUNCHER GUIDING CATHETER, PROWLER SELECT PLUS MICROCATHETER, AND NONCORDIS ECHELON MICROCATHETER WERE USED DURING THE PROCEDURE. THERE IS NO FURTHER PROCEDURAL INFORMATION AVAILABLE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY REPORTED PROBLEMS. THERE WERE NO REPORTED PRODUCT ISSUES, PROCEDURAL COMPLICATIONS OR ADVERSE EVENTS DURING THE PROCEDURE. THE PATIENT HAD NO NEUROLOGICAL DEFICIT UPON LEAVING THE ANGIOGRAPHY SUITE. THE DAY AFTER THE PROCEDURE, THE PATIENT WAS FOUND TO HAVE TWO SPOTS OF PUNCTUATE HIGH INTENSITY AREAS (HIA) IN THE LEFT OCCIPITAL LOBE. IT WAS REPORTED AS AN ASYMPTOMATIC CEREBRAL INFARCTION. THE ENTERPRISE VRD REMAINS IMPLANTED AND THE PROWLER MICROCATHETER IS NOT AVAILABLE FOR ANALYSIS. (B)(4): THE LOT NUMBER OF THE PROWLER SELECT PLUS MICROCATHETER IS NOT KNOWN; THEREFORE A DEVICE HISTORY RECORD REVIEW CANNOT BE COMPLETED. CEREBRAL INFARCTION (STROKE) IS A KNOWN POTENTIAL ADVERSE EVENT ASSOCIATED WITH THIS TYPE OF PROCEDURE AS OUTLINED IN THE INSTRUCTIONS FOR USE. BASED ON THE REPORTED INFORMATION IT IS NOT POSSIBLE TO DRAW A CONCLUSION REGARDING THE CLINICAL RELATIONSHIP BETWEEN THE DEVICES AND THE EVENT. PATIENT, PROCEDURAL, AND PHARMACOLOGICAL FACTORS MAY HAVE CONTRIBUTED TO THE EVENT. THERE IS NO INDICATION OF A RELATIONSHIP TO ANY DEVICE DESIGN, PERFORMANCE, OR MANUFACTURING ISSUES. THEREFORE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. THIS IS ONE OF TWO PRODUCTS USED DURING THE SAME PROCEDURE. PLEASE REFERENCE MFR. REPORT # 1058196-2010-00357 AND # 1058196-2010-00358.

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCTS: CHIKAI GUIDEWIRE, LAUNCHER GUIDING CATHETER, ECHELON, ENTERPRISE VRD STENT. THE PRODUCT IS NOT AVAILABLE FOR EVALUATION AND TESTING. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS UPON RECEIPT. THIS IS ONE OF TWO PRODUCTS USED DURING THE SAME PROCEDURE. PLEASE REFERENCE MFR REPORT # 1058196-2010-00358.

Description of Event or Problem · 1

THE REPORT RECEIVED FROM THE AFFILIATE INDICATED THAT THE PATIENT WAS ENROLLED IN A (B)(4). THE INDEX PROCEDURE WAS FOR A COIL EMBOLIZATION OF A CEREBRAL ANEURYSM ASSISTED BY IMPLANTATION OF AN ENTERPRISE VRD STENT. NO TARGET LESION INFORMATION WAS AVAILABLE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY REPORTED PROBLEMS. THERE WERE NO REPORTED PRODUCT ISSUES, PROCEDURAL COMPLICATIONS OR ADVERSE EVENTS DURING THE PROCEDURE. THE PATIENT HAD NO NEUROLOGICAL DEFICIT UPON LEAVING THE ANGIOGRAPHY SUITE. THE DAY AFTER THE PROCEDURE, THE PATIENT WAS REPORTED TO HAVE EXPERIENCED AN ASYMPTOMATIC CEREBRAL INFARCTION. THE PATIENT WAS FOUND TO HAVE TWO SPOTS OF HIGH INTENSITY AREAS (PUNCTUATE HIA) IN THE LEFT OCCIPITAL LOBE. THE PATIENT WAS ASYMPTOMATIC AND NO TREATMENT WAS PERFORMED. THE PHYSICIAN COMMENTED THAT IT WAS NOT KNOWN IF THE EVENT WAS CAUSED BY THE CATHETER MANIPULATION OR THE ENTERPRISE VRD DEVICE. THE PATIENT WAS REPORTED TO BE CURRENTLY IN STABLE CONDITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROWLER SELECT MICROCATHETERS CES MICROCATHETERS (KRA) KRA CORDIS NEUROVASCULAR, INC. UNK

Patients

Seq Age Sex Outcome Treatment
1 55 YR