FDA Adverse Event Malfunction Summary report: N

LEGEND FOOTED ATTACHMENT

MDR report key: 1915167 · Received December 6, 2010

Report

Report Number
1625507-2010-00075
Event Type
Malfunction
Date Received
December 6, 2010
Date of Event
November 1, 2010
Report Date
November 24, 2010
Manufacturer
MDT POWERED SURGICAL SOLUTIONS
Product Code
HBB
PMA / PMN Number
K020069
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

REPORT CONFIRMED. EVALUATION DETERMINED THAT THE FOOTED PORTION WAS DAMAGED BY TOOL CONTACT. IT WAS ALSO NOTED THAT SIGNS OF INTERNAL CORROSION WERE EVIDENT. ON FOLLOW-UP IT WAS CONFIRMED THAT THERE WAS NO PATIENT IMPACT. THE USER MANUAL CONTAINS THE FOLLOWING WARNING "THE ATTACHMENT SHOULD NOT BE USED IF ANY PART OF THE ATTACHMENT APPEARS TO BE BENT, LOOSE, MISSING, OR DAMAGED. EXCESSIVE PRESSURE OR IMPROPER HANDLING, SUCH AS BENDING OR PRYING, OF THE ATTACHMENT OR DISSECTING TOOL MAY CAUSE INJURY TO THE PATIENT, OPERATOR AND/OR OPERATING ROOM STAFF."

Description of Event or Problem · 1

REPAIR REQUEST INITIATED AND ESCALATED TO COMPLAINT FOR THE DEVICE WITH A REPORT THAT THE ATTACHMENT FOOT WAS CUT BY A TOOL. NO PATIENT IMPACT REPORTED. ON FOLLOW-UP, IT WAS NOTED THAT THIS WAS IDENTIFIED DURING A PROCEDURE AND IT WAS CONFIRMED THAT THERE WAS NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LEGEND FOOTED ATTACHMENT MOTOR, DRILL, PNEUMATIC HBB MDT POWERED SURGICAL SOLUTIONS N/A N/A

Patients

Seq Age Sex Outcome Treatment
1