HOMECHOICE CYCLER-REFURBISHED
Report
- Report Number
- 1423500-2010-06558
- Event Type
- Malfunction
- Date Received
- December 6, 2010
- Date of Event
- December 1, 2010
- Report Date
- December 1, 2010
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAS NOT YET BEEN RECEIVED BUT AN EVALUATION IS EXPECTED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE OFF CYCLER EXCHANGE. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER TO REQUEST ASSISTANCE ON THE HOME CHOICE (HC). PER THE INITIAL REPORT, THE HOME PATIENT (HP) STATED HE FELT SOME DISCOMFORT AND DID A MANUAL DRAIN, DRAIN VOLUME (DV) 4800ML. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASKED THE HP IF HE CALLED BAXTER OR THE REGISTERED NURSE (RN) ABOUT THE ISSUE AND THE HP STATED NO. THE TSR ASKED THE HP HOW HE FELT AND THE HP STATED FINE. THE HP STATED THE HC DISPLAYED DWELL 1 OF 4. THE TSR EXPLAINED TO THE HP THAT THE HC WOULD BE SWAPPED AND RECOMMENDED THAT HP CONTACT THE RN ABOUT HOW HE WAS FEELING AND THE HC BEING SWAPPED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. ON (B)(6) 2010 PRODUCT SURVEILLANCE CONTACTED THE HP PERITONEAL DIALYSIS NURSE (PDN) REGARDING THE REPORT OF HIGH DRAIN. THE PDN VERIFIED THE HP'S LARGEST PRESCRIBED FILL VOLUME (LPFV) IS 2500ML. THE PDN STATED THE HP HAS NOT REPORTED ANY SYMPTOMS OR OTHER DRAIN VOLUMES THAT MEET INCREASED INTRAPERITONEAL VOLUME (IIPV) CRITERIA. THE PDN CONFIRMED THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT AND THAT THE HP WAS DOING WELL WITH THE FOLLOWING TREATMENTS. NO ALLEGATIONS WERE MADE AGAINST ANY OF THE HP'S DIALYSIS PRODUCTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE CYCLER-REFURBISHED | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |