FDA Adverse Event Injury Summary report: N

PICOSURE

MDR report key: 19148796 · Received April 19, 2024

Report

Report Number
1222993-2024-00017
Event Type
Injury
Date Received
April 19, 2024
Report Date
April 19, 2024
Manufacturer
CYNOSURE LLC
Product Code
GEX
PMA / PMN Number
K210226
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CYNOSURE'S CLINICAL TEAM INVESTIGATED THE EVENT AND CONFIRMED USER ERROR WAS INVOLVED. PHOTOS OF THE PATIENT WAS PROVIDED AND CLINICAL DETERMINED PATIENT WAS INCORRECTLY ASSESSED. PATIENT'S FITZPATRICK APPEARS TO BE FITZ IV BUT WAS ASSESSED AS FITZ VI. SITE DID NOT PERFORM A TEST SPOT PRIOR TO TREATMENT ALTHOUGH RECOMMENDED PER LABELING. LABELING ALSO INSTRUCTS: LARGER SPOT SIZES WITH LOWER FLUENCE SHOULD BE USED IN HIGHLY PIGMENTED TATTOOS, AS WELL AS WITH TATTOOS ON EXTREMITIES. SITE PERFORMED TREATMENT WITH 4MM TIP ON THE EXTREMITY OF PATIENT'S WRIST AND INK WAS DENSE AND BLACK. CYNOSURE MEDICAL DIRECTOR ALSO REVIEWED THE INCIDENT AND SUSPECTS SPOT SIZE WAS TOO SMALL GIVEN THE DENSE BLACK INK AND THE FITZPATRICK OF THE PATIENT. PER CYNOSURE MEDICAL DIRECTOR: PATIENT'S WRIST HAS PARTIAL THICKNESS BURN AND FLANK FULL THICKNESS BURNS. SOME LEVEL OF SCARRING IS EXPECTED. PATIENT IS ADVISED TO FOLLOW UP WITH SOMEONE SKILLED WITH BURN TREATMENT. PATIENT WAS GIVEN TOPICAL SILVADENE AS PREVENTATIVE MEDICATION. THE DEVICE WAS EVALUATED AND FOUND TO BE OPERATING AS INTENDED WITHIN SPECIFICATION. COMPLIMENTARY VIRTUAL TRAINING WAS DISPATCHED TO THE SITE. BURNS ARE EXPECTED SIDE EFFECTS FROM LASER TREATMENTS, HOWEVER PERMANENT SCARRING IS A LIKELY OUTCOME AND THEREFORE REPORTABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT PATIENT EXPERIENCED A BURN FOLLOWING A TATTOO REMOVAL TREATMENT USING PICOSURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1111020 PICOSURE PICOSURE GEX CYNOSURE LLC

Patients

Seq Age Sex Outcome Treatment
1 NA Female Other