FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1914786 · Received December 4, 2010

Report

Report Number
1423500-2010-06522
Event Type
Malfunction
Date Received
December 4, 2010
Date of Event
November 8, 2010
Report Date
November 8, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT IS FOR A REPORT OF A CHECK SUPPLY LINE. THE USED SAMPLE WAS NOT RETURNED TO BAXTER FOR EVALUATION. PER THE REPORT, THE PATIENT NOTICED THAT THE SUPPLY BAG WAS NOT COMPLETELY SPIKED. FROM THE DATA WITHIN THE REPORT, THE ROOT CAUSE WAS USER ERROR. THIS REVIEW FOUND THE LABELING ADEQUATE FOR THE USER ERROR IN THE COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE LOT NUMBER WAS NOT PROVIDED; THEREFORE A BATCH REVIEW CANNOT BE CONDUCTED. THE SAMPLE IS NOT AVAILABLE AT THIS TIME. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

A CARE GIVER (CG) CONTACTED GLOBAL TECHNICAL SERVICES REGARDING A CHECK SUPPLY LINE ALARM THAT OCCURRED ON THE HOME CHOICE (HC) UNIT DURING PRIME. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) HAD THE CG CHECK THE SUPPLY LINE FOR KINKS/CLOSED CLAMPS; THE CG FOUND THE SPIKE WAS NOT PUSHED IN ALL THE WAY. THE CG PUSHED IN THE SPIKE THIS CORRECTED THE ALARM AND THE PRIME RESUMED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 72 YR