FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL

MDR report key: 1914754 · Received December 3, 2010

Report

Report Number
6000001-2010-05506
Event Type
Malfunction
Date Received
December 3, 2010
Date of Event
October 1, 2010
Report Date
October 15, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON FURTHER REVIEW OF THE EVENT HISTORY, THE OCCURRENCE DATE OF THIS EVENT WAS DETERMINED TO BE (B)(6) 2010. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. DURING PREVIOUS SERVICE, THE PUMP HEAD MODULE WAS REPLACED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THIS DEVICE WAS RETURNED TO BAXTER FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED. DEVICE EVALUATION CONFIRMED THE REPORTED CONDITION AS FAILURE CODE 808:03. THE ROOT CAUSE COULD NOT BE DETERMINED AT THIS TIME. NO REPAIRS WERE PERFORMED AT THIS TIME, AS THIS IS A BAXTER OWNED DEVICE. REVIEW OF THE DEVICE EVENT HISTORY DETERMINED THAT THE REPORTED CONDITION OF OCCURRED ON (B)(6) 2010 AND NOT THE CUSTOMER REPORTED OCCURRENCE DATE OF (B)(6) 2010. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 808:03. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. THE FACILITY REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. BAXTER?S REVIEW OF THE DEVICE EVENT HISTORY DETERMINED THE REPORTED CONDITION INTERRUPTED DELIVERY. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS 5.09.90, WHICH IS CLASSIFIED AS REMEDIATED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.