NEEDLE CLIPPING DEVICE SAFE CLIP
Report
- Report Number
- 3022699058-2024-00001
- Event Type
- Malfunction
- Date Received
- April 19, 2024
- Report Date
- July 12, 2024
- Manufacturer
- EMBECTA MEDICAL II LLC - PARSIPPANY, NJ
- Product Code
- FMI
- UDI-DI
- 00382903282357
- PMA / PMN Number
- K943683
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
CORRECTION TO: H6 (COMPONENT CODE, TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS) INVESTIGATION SUMMARY: SAMPLES WERE RECEIVED AND AN INVESTIGATION WAS PERFORMED. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. EMBECTA WAS NOT ABLE TO DUPLICATE OR CONFIRM THE INDICATED ISSUE AND BASED ON TREND ANALYSIS NO FURTHER ACTION IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED.
CONSUMER REPORTED THAT HER SAFE CLIP NEEDLE CLIPPER IS NOT WORKING, STATED THAT THE NEEDLE WILL NOT GO INTO THE SAFE CLIP. CONSUMER STATED THAT SHE HAD THE SAFE CLIP FOR AWHILE BUT IT IS NOT FULL BUT IT JUST STOPPED WORKING. LOT #: 1308206. CATALOG #: 328235. DATE OF EVENT: UNKNOWN. SAMPLES: AVAILABLE - SENDING MAIL KIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1686883 | NEEDLE CLIPPING DEVICE SAFE CLIP | NEEDLE, HYPODERMIC, SINGLE LUMEN | FMI | EMBECTA MEDICAL II LLC - PARSIPPANY, NJ | 328235 | 1308206 | 00382903282357 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |