FDA Adverse Event Malfunction Summary report: N

UNICEL® DXI 800 ACCESS® IMMUNOASSAY ANALYZER

MDR report key: 1914702 · Received December 3, 2010

Report

Report Number
2122870-2010-00856
Event Type
Malfunction
Date Received
December 3, 2010
Date of Event
November 4, 2010
Report Date
November 4, 2010
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K023764
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL TBHCG SAMPLES ARE RUN IN DUPLICATE. QC WAS WITHIN THE CUSTOMER'S ESTABLISHED RANGES DURING THE EVENT. SERVICE WAS DISPATCHED ON (B)(4) 2010. THE FIELD SERVICE ENGINEER (FSE) PERFORMED A PRECISION TEST, AND THE RESULTS MET PRECISION SPECIFICATIONS. NO CLEAR ROOT CAUSE FOR THE EVENT HAS BEEN DETERMINED TO DATE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO A LOWER THAN EXPECTED TOTAL BETA HCG RESULT GENERATED BY UNICEL DXI 800 ACCESS IMMUNOASSAY ANALYZER FOR ONE PATIENT. THE RESULT WAS NOT REPORTED OUT OF THE LABORATORY. SUBSEQUENT TESTING PRODUCED A HIGHER RESULT WHICH BETTER MATCHED THE PATIENT'S CLINICAL PICTURE. THE CUSTOMER DID NOT RECEIVE ANY REPORT OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXI 800 ACCESS® IMMUNOASSAY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER JJE BECKMAN COULTER, INC. DXI 800 NA

Patients

Seq Age Sex Outcome Treatment
1