SYNCHRON® CX5 DELTA CHEMISTRY ANALYZER
MDR report key: 1914676
·
Received December 3, 2010
Report
- Report Number
- 2050012-2010-01506
- Event Type
- Malfunction
- Date Received
- December 3, 2010
- Date of Event
- November 5, 2010
- Report Date
- November 5, 2010
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JGS
- PMA / PMN Number
- K188495
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
QC PERFORMANCE AT THE TIME OF THE EVENT IS UNKNOWN. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED, DECONTAMINATED THE SYSTEM, AND REPLACED SOME PARTS.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING COMPLAINTS OF SEVERAL LOW SODIUM (NA) RESULTS BEING GENERATED BY THE SYNCHRON CX5 DELTA CHEMISTRY ANALYZER, WHICH WERE REPORTED OUT OF THE LABORATORY. THE CUSTOMER STATED THAT THEY ROUTINELY RUN ANY SAMPLE THAT GIVES A NA RESULT LESS THAN 134 MMOL/L. OF THOSE SAMPLES THAT WERE RERUN THEY GENERALLY REPEAT 5-10MMOL/L HIGHER. AN EXAMPLE WAS PROVIDED OF A 131 MMOL/L NA REPEATING AT 140MMOL/L. PATIENT TREATMENT WAS NOT AFFECTED IN THIS EVENT. THE PHYSICIANS QUESTION THE RESULTS AND HAD THEM REPEATED PRIOR TO ADJUSTING TREATMENT.