SYNCHRON® CX5 DELTA CHEMISTRY ANALYZER

MDR report key: 1914676 · Received December 3, 2010

Report

Report Number
2050012-2010-01506
Event Type
Malfunction
Date Received
December 3, 2010
Date of Event
November 5, 2010
Report Date
November 5, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
JGS
PMA / PMN Number
K188495
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

QC PERFORMANCE AT THE TIME OF THE EVENT IS UNKNOWN. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED, DECONTAMINATED THE SYSTEM, AND REPLACED SOME PARTS.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING COMPLAINTS OF SEVERAL LOW SODIUM (NA) RESULTS BEING GENERATED BY THE SYNCHRON CX5 DELTA CHEMISTRY ANALYZER, WHICH WERE REPORTED OUT OF THE LABORATORY. THE CUSTOMER STATED THAT THEY ROUTINELY RUN ANY SAMPLE THAT GIVES A NA RESULT LESS THAN 134 MMOL/L. OF THOSE SAMPLES THAT WERE RERUN THEY GENERALLY REPEAT 5-10MMOL/L HIGHER. AN EXAMPLE WAS PROVIDED OF A 131 MMOL/L NA REPEATING AT 140MMOL/L. PATIENT TREATMENT WAS NOT AFFECTED IN THIS EVENT. THE PHYSICIANS QUESTION THE RESULTS AND HAD THEM REPEATED PRIOR TO ADJUSTING TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHRON® CX5 DELTA CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JGS BECKMAN COULTER INC. CX5 DELTA N/A