FDA Adverse Event Malfunction Summary report: N

BASAL/BOLUS INFUSOR, 12 PACK0.5ML/H X 1H

MDR report key: 1914608 · Received December 3, 2010

Report

Report Number
6000001-2010-05456
Event Type
Malfunction
Date Received
December 3, 2010
Date of Event
October 27, 2010
Report Date
November 11, 2010
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION AND/OR SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. A VISUAL EXAMINATION AND FUNCTIONAL TESTS WERE PERFORMED. DEVICE EVALUATION COULD NOT CONFIRM THE REPORTED CONDITION OF AN OVERINFUSION. THE ROOT CAUSE COULD NOT BE DETERMINED. THIS DEVICE IS A SINGLE USE DEVICE AND WAS DISCARDED. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. A FOLLOW UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT ONE (1) BASAL/BOLUS INFUSOR 0.5ML/HR DEVICE HAD OVERINFUSED DURING PATIENT USE AT THE HOSPITAL. ACCORDING TO THE REPORT, THE DEVICE WAS FILLED WITH 24ML OF SANDSTATINE (6ML OF THE DRUG AND 18ML OF NORMAL SALINE). THE ACTUAL FLOW RATE WAS 24ML IN 21HOURS. THERE IS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BASAL/BOLUS INFUSOR, 12 PACK0.5ML/H X 1H PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 09M004

Patients

Seq Age Sex Outcome Treatment
1